Validation of a high performance liquid chromatographic method for alpha amanitin determination in urine

被引:52
作者
Defendenti, C [1 ]
Bonacina, E [1 ]
Mauroni, M [1 ]
Gelosa, L [1 ]
机构
[1] Presidio Multizonale Igiene & Prevenz, Unita Riferimento Reg Micol, I-20129 Milan, Italy
关键词
column liquid chromatography; electrochemical detection; alpha amanitin;
D O I
10.1016/S0379-0738(98)00006-1
中图分类号
DF [法律]; D9 [法律]; R [医药、卫生];
学科分类号
0301 ; 10 ;
摘要
The objective of the present study was to develop and validate a liquid chromatographic method with electrochemical detection to measure alpha amanitin concentrations in urine after sample pretreatment with double mechanism (reversed phase/cation exchange) solid-phase extraction cartridges. The urine samples (10 mi) were purified and concentrated to 1 mi with elimination of matrix contaminants. The extracts were then separated by isocratic reversed-phase chromatography using a C18 column (4.6 mm x 25 cm) with a mobile phase composed of 0.005 M phosphate buffer (pH 7.2) and acetonitrile (90:10). Coulometric detection was performed by applying an oxidation potential of +500 mV to a porous graphite electrode in a low-volume analytical cell. The limit of quantitation was 10 ng/ml with a signal-to-noise ratio=25. The linearity studied on spiked urine was satisfactory (r=0.9966) from 10 ng/ml to 200 ng/ml. The average extraction recovery of alpha amanitin was 78%, determined using spiked urine samples ranging from 10-300 ng/ml. The intra-assay precision was checked at 10, 50 and 100 ng/ml levels (n=10) in spiked urine samples, with resulting coefficients of variation of 3.6%, 2% and 1.5%, respectively. (C) 1998 Elsevier Science Ireland Ltd.
引用
收藏
页码:59 / 68
页数:10
相关论文
共 6 条
[1]   DETERMINATION OF ALPHA-AMANITIN IN SERUM BY HPLC [J].
BELLIARDO, F ;
MASSANO, G ;
ANTIVELENI, C .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1983, 6 (03) :551-558
[2]  
BUSI C, 1977, NOUV PRESSE MED, V6, P2855
[3]   DIRECT, SIMULTANEOUS DETERMINATION OF ALPHA-AMANITIN, BETA-AMANITIN AND PHALLOIDINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
CACCIALANZA, G ;
GANDINI, C ;
PONCI, R .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1985, 3 (02) :179-185
[4]   DETERMINATION OF ALPHA-AMANITIN AND BETA-AMANITIN IN HUMAN BIOLOGICAL-FLUIDS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
JEHL, F ;
GALLION, C ;
BIRCKEL, P ;
JAEGER, A ;
FLESCH, F ;
MINCK, R .
ANALYTICAL BIOCHEMISTRY, 1985, 149 (01) :35-42
[5]   DETERMINATION OF ALPHA-AMANITIN BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
PASTORELLO, L ;
TOLENTINO, D ;
DALTERIO, M ;
PALADINO, R ;
FRIGERIO, A ;
BERGAMO, N ;
VALLI, A .
JOURNAL OF CHROMATOGRAPHY, 1982, 233 (DEC) :398-403
[6]   IMPROVED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION WITH AMPEROMETRIC DETECTION OF ALPHA-AMANITIN IN HUMAN PLASMA BASED ON ITS VOLTAMMETRIC STUDY [J].
TAGLIARO, F ;
SCHIAVON, G ;
BONTEMPELLI, G ;
CARLI, G ;
MARIGO, M .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1991, 563 (02) :299-311