Barriers to quantitative electron probe x-ray microanalysis for low voltage scanning electron microscopy

被引:40
作者
Newbury, DE [1 ]
机构
[1] Natl Inst Stand & Technol, Gaithersburg, MD 20899 USA
来源
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY | 2002年 / 107卷 / 06期
关键词
electron probe x-ray microanalyzer; energy dispersive spectrometry; low voltage microanalysis; scanning electron microscope; wavelength dispersive spectrometry; x-ray spectrometry;
D O I
10.6028/jres.107.049
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Low voltage x-ray microanalysis, defined as being performed with an incident beam energy less than or equal to5 keV, can achieve spatial resolution, laterally and in depth, of 100 nm or less, depending on the exact selection of beam energy and the composition of the target. The shallow depth of beam penetration, with the consequent short path length for x-ray absorption, and the low overvoltage, the ratio of beam energy to the critical ionization energy, both contribute to minimizing the matrix effects in quantitative x-ray microanalysis when the unknown is compared to pure element standards. The low beam energy restricts the energy of the atomic shells that can be excited, forcing the analyst to choose unfamiliar shells/characteristic peaks. The low photon energy shells are subject to low fluorescence yield, so that the peak-to-continuum background is reduced, severely limiting detectability. The limited resolution of semiconductor energy dispersive spectrometry results in frequent peak interference situations and further exacerbates detection limits. Future improvements to the x-ray spectrometry limitations are possible with x-ray optics-augmented wavelength dispersive spectrometry and microcalorimeter energy dispersive spectrometry.
引用
收藏
页码:605 / 619
页数:15
相关论文
共 12 条
[1]  
AGNELLO R, 1997, MICROSC MICROANAL, V3, P889
[2]  
[Anonymous], MICROSC MICROANAL
[3]  
FIORI CE, 1991, MICROBEAM ANAL 1991, P327
[4]  
Goldstein JDN., 2002, SCANNING ELECT MICRO
[5]  
Joy DC, 1998, AIP CONF PROC, V449, P653
[6]   PENETRATION AND ENERGY-LOSS THEORY OF ELECTRONS IN SOLID TARGETS [J].
KANAYA, K ;
OKAYAMA, S .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1972, 5 (01) :43-&
[7]  
Lifshin EC, 1980, P 8 INT C XRAY OPT M, P141
[8]  
MCCARTHY JJ, 1995, XRAY SPECTROMETRY EL, P67
[9]   Lowering the limit of detection in high spatial resolution electron beam microanalysis with the microcalorimeter energy dispersive X-ray spectrometer [J].
Newbury, D ;
Wollman, D ;
Irwin, K ;
Hilton, G ;
Martinis, J .
ULTRAMICROSCOPY, 1999, 78 (1-4) :73-88
[10]   High-resolution, energy-dispersive microcalorimeter spectrometer for X-ray microanalysis [J].
Wollman, DA ;
Irwin, KD ;
Hilton, GC ;
Dulcie, LL ;
Newbury, DE ;
Martinis, JM .
JOURNAL OF MICROSCOPY-OXFORD, 1997, 188 :196-223