Analytical evaluation of an improved procedure for measuring thyrotropin

被引:2
作者
Gonzalez-De-La-Presa, B
Palacios, G
Bonnin, R
Martinez, JM
Navarro, MA
机构
[1] Ciutat Sanitaria & Univ Bellvitge, Inst Catala Salut, Secc Hormonas, E-08022 Barcelona, Spain
[2] Abbot Cient, Madrid, Spain
关键词
thyrotropin; functional sensitivity; imprecision; fluorescence-enhanced immunoenzymometric assay;
D O I
10.1515/CCLM.1998.019
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The analytical characteristics of the AxSYM Ultrasensitive hTSH-II (Abbott Laboratories) procedure for quantitation of serum thyrotropin (TSH) concentration were evaluated. Within-and between-run imprecisions, functional sensitivity, analytical interval and relative inaccuracy with respect to the Enzymum-Test TSH (Boehringer Mannheim) were studied. In all cases, the within-run and between-run of coefficients variation were lower than 6.69% and 8.12% respectively. The measurement range was tested with serial dilutions of a serum with a high thyrotropin concentration, and the procedure was found to be linear up to at least 87.0 mIU/l. The functional sensitivity was 0.018 mIU/l. The relative inaccuracy study (Passing-Bablok non-parametric linear regression) produced the following linear equation: (AxSYM) = 1.02. (ES-700) - 0.03 mIU/l, with 95% confidence intervals of a (-0.05; -0.01); b (0.98; 1.06).
引用
收藏
页码:103 / 106
页数:4
相关论文
共 9 条
[1]  
COTLOVE E, 1970, CLIN CHEM, V16, P1028
[2]  
FRASER CG, 1992, EUR J CLIN CHEM CLIN, V30, P311
[3]  
HARRIS EK, 1979, AM J CLIN PATHOL, V72, P374
[4]  
HAY ID, 1991, CLIN CHEM, V37, P2002
[5]  
PASSING H, 1983, J CLIN CHEM CLIN BIO, V21, P709
[6]   QUALITY GOALS FOR HORMONE TESTING [J].
RICOS, C ;
ARBOS, MA .
ANNALS OF CLINICAL BIOCHEMISTRY, 1990, 27 :353-358
[7]  
RODEN M, 1993, CLIN CHEM, V39, P544
[9]  
Wood WG, 1996, EUR J CLIN CHEM CLIN, V34, P151