Electrochemical evaluation of Ti-13Nb-13Zr, Ti-6Al-4V and Ti-6Al-7Nb alloys for biomedical application by long-term immersion tests

被引:144
作者
Assis, S. L. [1 ]
Costa, I. [1 ]
机构
[1] IPEN CNEN SP, Ctr Mat Sci & Technol, Sao Paulo, Brazil
来源
MATERIALS AND CORROSION-WERKSTOFFE UND KORROSION | 2007年 / 58卷 / 05期
关键词
D O I
10.1002/maco.200604027
中图分类号
T [工业技术];
学科分类号
08 [工学];
摘要
In this investigation the electrochemical behaviour of the Ti13Nb-13Zr, Ti-6Al-4V and Ti-6Al-7Nb alloys, for application as implant materials was evaluated in Hanks' solution by electrochemical techniques. The alloys were immersed in this solution for 410 days and periodically they were tested by electrochemical impedance spectroscopy. At the end of this period, polarization curves of the three titanium alloys were obtained. The electrochemical impedance experimental results were interpreted using an equivalent electrical circuit that simulates a duplex structure oxide composed of an inner compact layer, here called barrier layer, and an outer and porous layer. The results indicated that all the alloys present a very high corrosion resistance in the electrolyte used, typical of passive alloys, and that the corrosion resistance is mainly due to the barrier layer. The passive like behaviour was maintained during the whole period of test.
引用
收藏
页码:329 / 333
页数:5
相关论文
共 18 条
[1]
[Anonymous], 2006, ELECTROCHIM ACTA, DOI [DOI 10.1016/J.ELECTACTA.2005.02.121, DOI 10.1016/j.electacta.2005.02.121]
[2]
A comparative study of the in vitro corrosion behavior and cytotoxicity of a superferritic stainless steel, a Ti-13Nb-13Zr alloy, and an austenitic stainless steel in Hank's solution [J].
Assis, SL ;
Rogero, SO ;
Antunes, RA ;
Padilha, AF ;
Costa, I .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2005, 73B (01) :109-116
[3]
ASSIS SL, 2006, THESIS IPEN CNEN SP
[4]
Bundy K.J., 1995, CORROSION TESTS STAN, P411
[5]
USE OF AC IMPEDANCE METHODS TO STUDY THE CORROSION BEHAVIOR OF IMPLANT ALLOYS [J].
BUNDY, KJ ;
DILLARD, J ;
LUEDEMANN, R .
BIOMATERIALS, 1993, 14 (07) :529-536
[6]
Corrosion behaviour of titanium in biofluids containing H2O2 studied by electrochemical impedance spectroscopy [J].
Fonseca, C ;
Barbosa, MA .
CORROSION SCIENCE, 2001, 43 (03) :547-559
[7]
ELECTROCHEMISTRY AND PASSIVITY OF A TI-15MO-3NB-3AL BETA-TITANIUM ALLOY IN AMBIENT-TEMPERATURE AQUEOUS CHLORIDE SOLUTIONS [J].
KOLMAN, DG ;
SCULLY, JR .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1993, 140 (10) :2771-2779
[8]
ELECTROCHEMISTRY AND PASSIVITY OF TI-15 V-3 CR-3 AL-3 SN BETA-TITANIUM ALLOY IN AMBIENT-TEMPERATURE AQUEOUS CHLORIDE SOLUTIONS [J].
KOLMAN, DG ;
SCULLY, JR .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1994, 141 (10) :2633-2641
[9]
Electrochemical impedance spectroscopy characterization of passive film formed on implant Ti-6Al-7Nb alloy in Hank's solution [J].
Lavos-Valereto, IC ;
Wolynec, S ;
Ramires, I ;
Guastaldi, AC ;
Costa, I .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2004, 15 (01) :55-59
[10]
Passive film on orthopaedic TiAlV alloy formed in physiological solution investigated by X-ray photoelectron spectroscopy [J].
Milosev, I ;
Metikos-Hukovic, M ;
Strehblow, HH .
BIOMATERIALS, 2000, 21 (20) :2103-2113