摘要
We investigate the way data are used in the algorithm proposed by Kudo and Saito for the exact reconstruction of long objects from axially truncated cone-beam projections. Specifically, we show that the algorithm wastes a large part of the data. To overcome the problem we propose to use a vertex path consisting of two crossing ellipses, for which we devised a new reconstruction algorithm, called the cross algorithm, which does not waste data and is still suitable to exactly handle axial truncation. Results of reconstruction are presented on simulated data and real data from an experimental scanner.
引用
收藏
页码:797 / 810
页数:14
相关论文
共 22 条
  • [1] AXELSSONJACOBSO.C, 1996, LINKOPING STUDIES SC, V427
  • [2] CLACK R, 1994, P SOC PHOTO-OPT INS, V2299, P230, DOI 10.1117/12.179253
  • [3] A CONE-BEAM RECONSTRUCTION ALGORITHM USING SHIFT-VARIANT FILTERING AND CONE-BEAM BACKPROJECTION
    DEFRISE, M
    CLACK, R
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (01) : 186 - 195
  • [4] PRACTICAL CONE-BEAM ALGORITHM
    FELDKAMP, LA
    DAVIS, LC
    KRESS, JW
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06): : 612 - 619
  • [5] GRANGEAT P, 1991, LECT NOTES MATH, V1497, P66
  • [6] GRANGEAT P, 1997, STUDIES HLTH TECHNOL, V30, P343
  • [7] GRANGEAT P, 1997, ST HEAL T, V30, P29
  • [8] 3-DIMENSIONAL SPECT RECONSTRUCTION OF COMBINED CONE BEAM AND PARALLEL BEAM DATA
    JASZCZAK, RJ
    LI, JY
    WANG, H
    COLEMAN, RE
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (03) : 535 - 548
  • [9] DERIVATION AND IMPLEMENTATION OF A CONE-BEAM RECONSTRUCTION ALGORITHM FOR NONPLANAR ORBITS
    KUDO, H
    SAITO, T
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (01) : 196 - 211
  • [10] KUDO H, 1991, CONFERENCE RECORD OF THE 1991 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE, VOLS 1-3, P1958, DOI 10.1109/NSSMIC.1991.259259