Rapid serum vancomycin assay by high-performance liquid chromatography using a semipermeable surface packing material column

被引:25
作者
Furuta, I
Kitahashi, T
Kuroda, T
Nishio, H
Oka, C
Morishima, Y
机构
[1] Kinki Univ, Sch Med, Dept Clin Pathol, Osakasayama, Japan
[2] Kinki Univ Hosp, Cent Lab, Osakasayama, Japan
关键词
vancomycin; semipermeable surface; direct serum injection measurement; isocratic HPLC; MRSA; therapeutic drug monitoring (TDM);
D O I
10.1016/S0009-8981(00)00307-7
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
A new isocratic high-performance liquid chromatographic (HPLC) assay has been developed for vancomycin that uses direct injection of microquantities of serum into a separation column filled with octyl-C-8 silica support that has a semipermeable surface. A mixture of disodium hydrogen phosphate buffer (pH 7.0) and acetonitrile is used as the mobile phase, and vancomycin is directly detected at 240 nm. The minimum limit of detection was 0.5 mug/ml at a signal-to-noise ratio of 3:1. Linearity was established from 0 to 100 mug/ml The coefficient of variation for within-run reproducibility was 1.1-2.7% for a concentration range of 2.9-52.5 mug/ml, for day-to-day reproducibility it was 4.0% and 3.1% for a concentration range of 5.8-26.4 mug/ml, and the recovery rate was 94-105%. There was no interference from 41 antibiotics or other drugs currently in use. The correlation coefficient between the fluorescence polarization immunoassay (x) and this method (y) was 0.995 with a linens equation, y = 1.06x - 0.924. This method is. simple, rapid, and provides an economical quantification of serum vancomycin. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:31 / 39
页数:9
相关论文
共 13 条
[1]  
ANNE L, 1989, THER DRUG MONIT, V11, P585
[2]   Quantitation of vancomycin and its crystalline degradation product (CDP-1) in human serum by high performance liquid chromatography [J].
Backes, DW ;
Aboleneen, HI ;
Simpson, JA .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1998, 16 (08) :1281-1287
[3]   DETERMINATION OF VANCOMYCIN IN SERUM AND TISSUES BY COLUMN LIQUID-CHROMATOGRAPHY USING SOLID-PHASE EXTRACTION [J].
BAUCHET, J ;
PUSSARD, E ;
GARAUD, JJ .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1987, 414 (02) :472-476
[4]   Interference with vancomycin fluorescence polarisation immunoassay in a nonuraemic patient [J].
Bowhay, S ;
Timms, P .
THERAPEUTIC DRUG MONITORING, 1997, 19 (01) :117-119
[5]   THE SEMIPERMEABLE SURFACE, A NEW RESTRICTED ACCESS MEDIUM [J].
GLUNZ, LJ ;
PERRY, JA ;
INVERGO, B ;
WAGNER, H ;
SZCZERBA, TJ ;
RATEIKE, JD ;
GLUNZ, PW .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1992, 15 (08) :1361-1379
[6]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ANALYSIS OF VANCOMYCIN IN PLASMA, BONE, ATRIAL APPENDAGE TISSUE AND PERICARDIAL FLUID [J].
GREENE, SV ;
ABDALLA, T ;
MORGAN, SL .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1987, 417 (01) :121-128
[7]   RAPID AND SPECIFIC METHOD FOR THE DETERMINATION OF VANCOMYCIN IN PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY ON AN AMINOPROPYL COLUMN [J].
HOSOTSUBO, H .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1989, 487 (02) :421-427
[8]   MEASUREMENT OF VANCOMYCIN IN RENALLY IMPAIRED PATIENT SAMPLES USING A NEW HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY METHOD WITH VITAMIN-B12 INTERNAL STANDARD - COMPARISON OF HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY, EMIT, AND FLUORESCENCE POLARIZATION IMMUNOASSAY METHODS [J].
HU, MW ;
ANNE, L ;
FORNI, T ;
GOTTWALD, K .
THERAPEUTIC DRUG MONITORING, 1990, 12 (06) :562-569
[9]  
KATSUYAMA Y, 1998, JPN J THER DRUG MONI, V15, P334
[10]  
KITAHASHI T, 1997, JJCLA, V22, P205