Multi-frequency harmonic arrays: initial experience with a novel transducer concept for nonlinear contrast imaging

被引:15
作者
Forsberg, F
Shi, WT
Jadidian, B
Winder, AA
机构
[1] Thomas Jefferson Univ, Dept Radiol, Div Ultrasound, Philadelphia, PA 19107 USA
[2] Layered Mfg Inc, New Brunswick, NJ 08901 USA
[3] Acoust Sci Associates, Westport, CT 06880 USA
关键词
transducer design; transducer arrays; ultrasound contrast agent; harmonic imaging; subharmonic imaging;
D O I
10.1016/j.ultras.2004.04.005
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Nonlinear contrast imaging modes such as second harmonic imaging (HI) and subharmonic imaging (SHI) are increasingly important for clinical applications. However, the performance of currently available transducers for HI and SHI is significantly constrained by their limited bandwidth. To bypass this constraint, a novel transducer concept termed multi-frequency harmonic transducer arrays (MFHA's) has been designed and a preliminary evaluation has been conducted. The MFHA may ultimately be used for broadband contrast enhanced HI and SHI with high dynamic range and consists of three multi-element piezo-composite sub-arrays (A-C) constructed so the center frequencies are 4f(A) = 2f(B) = f(C) (specifically 2.5/5.0/10.0 MHz and 1.75/3.5/7.0 MHz). In principle this enables SHI by transmitting on sub-array C receiving on B and, similarly, from B to A as well as HI by transmitting on A receiving on B and, likewise, from B to C. Initially transmit and receive pressure levels of the arrays were measured with the elements of each sub-array wired in parallel. Following contrast administration, preliminary in vitro HI and SHI signal-to-noise ratios of up to 40 dB were obtained. In conclusion, initial design and in vitro characterization of two MFHA's have been performed. They have an overall broad frequency bandwidth of at least two octaves. Due to the special design of the array assembly, the SNR for HI and SHI was comparable to that of regular B-mode and better than commercially available HI systems. However, further research on multi-element MFHA's is required before their potential for in vivo nonlinear contrast imaging can be assessed. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:79 / 85
页数:7
相关论文
共 20 条
[1]   Piezoelectric ceramics and composites via rapid prototyping techniques [J].
Bandyopadhyay, A. ;
Panda, R. K. ;
McNulty, T. F. ;
Mohammadi, F. ;
Danforth, S. C. ;
Safari, A. .
RAPID PROTOTYPING JOURNAL, 1998, 4 (01) :37-49
[2]  
Bandyopadhyay A, 1997, J AM CERAM SOC, V80, P1366, DOI 10.1111/j.1151-2916.1997.tb02993.x
[3]   Subharmonic signal generation from contrast agents in simulated neovessels [J].
Bhagavatheeshwaran, G ;
Shi, WT ;
Forsberg, F ;
Shankar, PM .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2004, 30 (02) :199-203
[4]   Contrast superharmonic imaging: A feasibility study [J].
Bouakaz, A ;
Krenning, BJ ;
Vletter, WB ;
ten Cate, FJ ;
De Jong, N .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2003, 29 (04) :547-553
[5]   Super harmonic imaging: A new imaging technique for improved contrast detection [J].
Bouakaz, A ;
Frigstad, S ;
Ten Cate, FJ ;
de Jong, N .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2002, 28 (01) :59-68
[6]   Nondestructive subharmonic imaging [J].
Chomas, J ;
Dayton, P ;
May, D ;
Ferrara, K .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2002, 49 (07) :883-892
[7]   HIGHER HARMONICS OF VIBRATING GAS-FILLED MICROSPHERES .2. MEASUREMENTS [J].
DEJONG, N ;
CORNET, R ;
LANCEE, CT .
ULTRASONICS, 1994, 32 (06) :455-459
[8]   HIGHER HARMONICS OF VIBRATING GAS-FILLED MICROSPHERES .1. SIMULATIONS [J].
DEJONG, N ;
CORNET, R ;
LANCEE, CT .
ULTRASONICS, 1994, 32 (06) :447-453
[9]   On the feasibility of real time, in vivo harmonic imaging with proteinaceous microspheres [J].
Forsberg, F ;
Goldberg, BB ;
Liu, JB ;
Merton, DA ;
Rawool, NM .
JOURNAL OF ULTRASOUND IN MEDICINE, 1996, 15 (12) :853-860
[10]   Subharmonic imaging of contrast agents [J].
Forsberg, F ;
Shi, WT ;
Goldberg, BB .
ULTRASONICS, 2000, 38 (1-8) :93-98