Scanning electron microscope appearances of Lightspeed instruments used clinically: a pilot study

被引:29
作者
Marending, M [1 ]
Lutz, F [1 ]
Barbakow, F [1 ]
机构
[1] Univ Zurich, Sch Dent, Dept Prevent Dent Periodontol & Cariol, Div Endodontol, CH-8028 Zurich, Switzerland
关键词
instruments; Lightspeed; SEM;
D O I
10.1046/j.1365-2591.1998.t01-1-00104.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The cutting heads of the first 18 instruments in six sets of clinically used Lightspeed instruments (sizes 20-65) were examined in a scanning electron microscope (SEM), Three sets had been used by full-time faculty and discarded as follows: instrument sizes 20-32.5 after 18 treated root canals; instrument sizes 35-47.5 after 36 treated root canals; and instrument sizes 50-65 after 54 treated root canals, Three other sets had been used by a private clinician who discarded all Lightspeed instruments after preparing 20 root canals, The cutting heads of the 108 instruments were examined for the presence of surface debris, metal strips, metal flash, disrupted cutting edges, microfractures, tears, pitting and fretting and fatigue cracks, Fewer instruments ultrasonically cleaned before examination in the SEM had debris than those not cleaned before examination, The prevalence of metal strips and metal flash was similar for both groups, In contrast, microfractures and tears were more prevalent in instruments used by the full-time faculty group, who used their instruments more frequently before discarding them, Pitting and fretting occurred equally frequently in both groups whilst no fatigue cracks were noted in any of the 108 instruments, Overusage of Lightspeed instruments may predispose the flutes of the cutting heads to microfractures and clinicians should be aware of this possibility.
引用
收藏
页码:57 / 62
页数:6
相关论文
共 17 条
[1]   The 'Lightspeed' preparation technique evaluated by Swiss clinicians after attending continuing education courses [J].
Barbakow, F ;
Lutz, F .
INTERNATIONAL ENDODONTIC JOURNAL, 1997, 30 (01) :46-50
[2]   A COMPARISON OF ROOT-CANAL PREPARATIONS USING NI-TI HAND, NI-TI ENGINE-DRIVEN, AND K-FLEX ENDODONTIC INSTRUMENTS [J].
GLOSSON, CR ;
HALLER, RH ;
DOVE, SB ;
DELRIO, CE .
JOURNAL OF ENDODONTICS, 1995, 21 (03) :146-151
[3]   Assessing apical deformation and transportation following the use of LightSpeed(TM) root-canal instruments [J].
Knowles, KI ;
Ibarrola, JL ;
Christiansen, RK .
INTERNATIONAL ENDODONTIC JOURNAL, 1996, 29 (02) :113-117
[4]  
Lautenschlager E P, 1977, J Endod, V3, P175, DOI 10.1016/S0099-2399(77)80092-7
[5]  
*LIGHTSP TECHN INC, 1996, LIGHTSP REF GUID
[6]   Morphometric video analysis of the engine-driven nickel-titanium Lightspeed instrument system [J].
Marsicovetere, ES ;
Clement, DJ ;
delRio, CE .
JOURNAL OF ENDODONTICS, 1996, 22 (05) :231-235
[7]   FILE DAMAGE DURING ROOT-CANAL PREPARATION [J].
MONTGOMERY, S ;
CANALES, ML ;
DELRIO, CE .
JOURNAL OF ENDODONTICS, 1984, 10 (02) :45-47
[8]  
Paucksch E, 1993, ZERSPANTECHNIK
[9]   Cyclic fatigue testing of nickel-titanium endodontic instruments [J].
Pruett, JP ;
Clement, DJ ;
Carnes, DL .
JOURNAL OF ENDODONTICS, 1997, 23 (02) :77-85
[10]   CLOCKWISE OR COUNTERCLOCKWISE [J].
ROANE, JB ;
SABALA, C .