Less invasive stabilization of complex tibial plateau fractures -: A biomechanical evaluation of a unilateral locked screw plate and double plating

被引:64
作者
Gösling, T
Schandelmaier, P
Marti, A
Hufner, T
Partenheimer, A
Krettek, C
机构
[1] Hannover Med Sch, Unfallchirurg Klin, Trauma Dept, D-30625 Hannover, Germany
[2] Leopold Franzens Univ, Dept Trauma Surg, Innsbruck, Austria
[3] Robert Mathys Stiftung, Bettlach, Switzerland
关键词
tibial plateau; biomechanics; internal fixator; locked screw plate; plate;
D O I
暂无
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: To compare the vertical subsidence in a bicondylar tibial plateau fracture model stabilized either by a unilateral locked screw plate (LSP) or by double plating. Design: Biomechanical cadaver study. Intervention: A 41-C1 fracture model was created in eight pairs of fresh-frozen human cadaver tibiae. Stabilization was performed either by open reduction and internal fixation (ORIF) using a lateral L-buttress plate and a medial four-hole, one-third tubular antiglide-plate or by a lateral LSP. Four load levels (400N, 800N, 1200N, 1600N), each with five cycles, were consecutively applied to the medial plateau. Main Outcome Measurements: The vertical plastic deformation at the end of each cycle was the main parameter of interest. Statistical analysis was performed with the two-way ANOVA test for repeated measurements. Each individual loading level was analyzed separately using Student t test. Results: In one pair, both fixation techniques failed at the first loading cycle of 1200N. One ORIF fixation failed at the first loading cycle of 1600N. The average plastic vertical subsidence was 0.40 mm (LSP) and 0.25 mm (ORIF) at 400N (P = 0.291) 0.83 mm (LSP) and 0.81 mm (ORIF) at 800N (P = 8.82), 1.06 mm (LSP) and 0.96 min (ORIF) at 1200N (P = 0.98), and 1.54 mm (LSP) and 1.14 mm (ORIF) at 1600N (P = 0.53). Vertical subsidence depended on the applied load (P = 0.002), but not on the method of fixation (P = 0.236). Conclusion: Both fixation techniques have a high resistance to vertical subsidence even with loads exceeding the average body weight. No statistically significant difference was seen between the two methods of fixation.
引用
收藏
页码:546 / 551
页数:6
相关论文
共 24 条
[1]
Effect of articular step-off and meniscectomy on joint alignment and contact pressures for fractures of the lateral tibial plateau [J].
Bai, B ;
Kummer, FJ ;
Sala, DA ;
Koval, KJ ;
Wolinsky, PR .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2001, 15 (02) :101-106
[2]
Bicondylar fracture of the posterior aspect of the tibial plateau - A case report and a modified operative approach [J].
Carlson, DA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (07) :1049-1052
[3]
The development of the distal femur Less Invasive Stabilization System (LISS) [J].
Frigg, R ;
Appenzeller, A ;
Christensen, R ;
Frenk, A ;
Gilbert, S ;
Schavan, R .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2001, 32 :24-31
[4]
Frigg R, 2001, Injury, V32 Suppl 2, P63, DOI 10.1016/S0020-1383(01)00127-9
[5]
Osteosynthesis of tibial plateau fractures in the elderly - A matched-pair-analysis [J].
Gerich, T ;
Blauth, M ;
Witte, F ;
Krettek, C .
UNFALLCHIRURG, 2001, 104 (01) :50-56
[6]
Precision and intersite correlation of bone densitometry at the radius, tibia and femur with peripheral quantitative CT [J].
Groll, O ;
Lochmüller, EM ;
Bachmeier, M ;
Willnecker, J ;
Eckstein, F .
SKELETAL RADIOLOGY, 1999, 28 (12) :696-702
[7]
Tibial plateau fractures [J].
Hertel, P .
UNFALLCHIRURG, 1997, 100 (07) :508-523
[8]
INTERFRAGMENTAL COMPRESSION OF ZESPOL OSTEOSYNTHESIS SYSTEM - AN EXPLORATORY BIOMECHANIC EXPERIMENT [J].
HOPF, T ;
OSTHEGE, S .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1987, 125 (05) :546-552
[9]
A biomechanical analysis of internal fixation of complex tibial plateau fractures [J].
Horwitz, DS ;
Bachus, KN ;
Craig, MA ;
Peters, CL .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1999, 13 (08) :545-549
[10]
Treatment of complex (Schatzker Type VI) fractures of the tibial plateau with circular wire external fixation: Retrospective case review [J].
Kumar, A ;
Whittle, AP .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2000, 14 (05) :339-344