Blood flow around hollow fibers

被引:21
作者
Dierickx, PWT [1 ]
De Wachter, D [1 ]
Verdonck, PR [1 ]
机构
[1] Univ Ghent, IBITECH, Hydraul Lab, B-9000 Ghent, Belgium
关键词
oxygenator; membrane; non-Newtonian; crossflow; resistance;
D O I
10.1177/039139880002300903
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In an artificial lung, blood is oxygenated and flows around a bundle of hollow fibers while gas flows inside the fiber. The objective of this study is to understand the hydrodynamics of three different fiber banks (inline square IS, staggered square SS and equilateral triangle ET) and;to investigate the influence of both a Newtonian and non-Newtonian Casson viscosity model on the flow field. A two-dimensional finite element model for permanent, isothermal, laminar blood now perpendicular to hollow fibers is used. All fibers are assumed identical, straight and parallel. Porosity ranges from 0.4 to 0.6 and Reynolds number varies from I to 60. For a given Re, ET generates less resistance than SS, the latter being comparable with IS. A lower porosity increases resistance. Non-Newtonian viscosity affects velocity patterns only at low Re (<0.5) and higher porosity (>0.5). Resistance at low Re is significantly elevated in the fiber banks due to an overall increase in viscosity This model makes it possible to study the influence of geometry and viscosity on hydrodynamics in fiber banks and may aid in the optimization of hollow fiber artificial lung design.
引用
收藏
页码:610 / 617
页数:8
相关论文
共 20 条
[1]   COMPUTATIONALLY 2-DIMENSIONAL FINITE-DIFFERENCE MODEL FOR HOLLOW-FIBER BLOOD-GAS EXCHANGE DEVICES [J].
BAKER, DA ;
HOLTE, JE ;
PATANKAR, SV .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1991, 29 (05) :482-488
[2]  
Bergelin O.P., 1952, Trans. ASME, P953
[3]  
BERGELIN OP, 1950, T ASME, V72, P881
[4]  
BERGELIN OP, 1954, T ASME, V76, P841
[5]  
Bird R.B., 2006, TRANSPORT PHENOMENA, Vsecond, DOI 10.1002/aic.690070245
[6]   BLOOD-FLOW OUTSIDE REGULARLY SPACED HOLLOW FIBERS - THE FUTURE CONCEPT OF MEMBRANE DEVICES [J].
CATAPANO, G ;
WODETZKI, A ;
BAURMEISTER, U .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1992, 15 (06) :327-330
[7]   Arteriovenous extracorporeal carbon dioxide removal - A mathematical model and experimental evaluation [J].
Conrad, SA ;
Brown, EG ;
Grier, LR ;
Baier, J ;
Blount, J ;
Heming, T ;
Zwischenberger, JB ;
Bidani, A .
ASAIO JOURNAL, 1998, 44 (04) :267-277
[8]  
Cuvelier C, 1986, FINITE ELEMENT METHO
[9]  
DIERICKX P, IN PRESS ASAIO J
[10]   THE INFLUENCE OF REYNOLDS-NUMBER UPON THE APPARENT PERMEABILITY OF SPATIALLY PERIODIC ARRAYS OF CYLINDERS [J].
EDWARDS, DA ;
SHAPIRO, M ;
BARYOSEPH, P ;
SHAPIRA, M .
PHYSICS OF FLUIDS A-FLUID DYNAMICS, 1990, 2 (01) :45-53