Enhanced semiquantitative multi-analysis of trace elements in environmental samples using inductively coupled plasma mass spectrometry

被引:26
作者
Bayon, MM [1 ]
Alonso, JIG [1 ]
Medel, AS [1 ]
机构
[1] Univ Oviedo, Fac Chem, Dept Phys & Analyt Chem, E-33006 Oviedo, Spain
关键词
inductively coupled plasma mass spectrometry; semiquantitative analysis; trace element; matrix effect; natural water; sediment;
D O I
10.1039/a707496d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A semiquantitative procedure for the determination of trace metals in natural waters and sediments by ICP-MS was developed, The method is based on the establishment of the instrumental molar response curve versus mass using a multielemental solution containing elements from Mg to Ri with different ionisation potentials, The ionisation conditions in the plasma (electron density, n(e), and ionisation temperature, T-ion) mere determined by least-squares fitting, using the Saha equation, by adjusting the molar response curve to a third-order polynomial. The effect of matrix constituents, such as Na+, K+, Ca2+, Mg2+, Cl- and SO42-, on both the shape of the molar response curve and on the plasma ionisation conditions was evaluated, It was observed that the ionisation degrees calculated in the plasma did not change significantly in the presence of the matrix constituents, However, the shape of the molar response curve was affected by high concentrations of matrix components and this effect was corrected by adding internal standards to the samples, The semiquantitative procedure was applied to the analysis of local natural waters and the results were compared with those obtained by external calibration and standard additions, Validation was performed by analysis of riverine water (SLRS-3) and sediment (PACS-1 and MESS-2) reference materials from the National Research Council of Canada with satisfactory result.
引用
收藏
页码:277 / 282
页数:6
相关论文
共 20 条
[1]   PERFORMANCE-CHARACTERISTICS OF A GLOVE BOX INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETER FOR THE ANALYSIS OF NUCLEAR-MATERIALS [J].
ALONSO, JIG ;
THOBYSCHULTZENDORFF, D ;
GIOVANONNE, B ;
KOCH, L ;
WIESMANN, H .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (05) :673-679
[2]   DETERMINATION OF FISSION-PRODUCTS AND ACTINIDES IN SPENT NUCLEAR-FUELS BY ISOTOPE-DILUTION ION CHROMATOGRAPHY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
ALONSO, JIG ;
SENA, F ;
ARBORE, P ;
BETTI, M ;
KOCH, L .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1995, 10 (05) :381-393
[3]   SEMIQUANTITATIVE ANALYSIS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
AMARASIRIWARDENA, CJ ;
GERCKEN, B ;
ARGENTINE, MD ;
BARNES, RM .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1990, 5 (06) :457-462
[4]   Semiquantitative analysis of biological materials by inductively coupled plasma-mass spectrometry [J].
Amarasiriwardena, D ;
Durrant, SF ;
Lasztity, A ;
Krushevska, A ;
Argentine, MD ;
Barnes, RM .
MICROCHEMICAL JOURNAL, 1997, 56 (03) :352-372
[5]   STUDY OF THE EFFECTS OF CONCOMITANT ELEMENTS IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BEAUCHEMIN, D ;
MCLAREN, JW ;
BERMAN, SS .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1987, 42 (03) :467-490
[6]   Determination of mercury species in sea-water by liquid chromatography with inductively coupled plasma mass spectrometric detection [J].
Bloxham, MJ ;
Gachanja, A ;
Hill, SJ ;
Worsfold, PJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (02) :145-148
[7]   SEMIQUANTITATIVE ANALYSIS WITH LASER-ABLATION INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
CROMWELL, EF ;
ARROWSMITH, P .
ANALYTICAL CHEMISTRY, 1995, 67 (01) :131-138
[8]   ELECTRONIC PARTITION FUNCTIONS OF ATOMS AND IONS BETWEEN 1500 DEGREES K AND 7000 DEGREES K [J].
DEGALAN, L ;
SMITH, R ;
WINEFORDNER, JD .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1968, B 23 (08) :521-+
[9]   SEMIQUANTITATIVE ANALYSIS OF ENVIRONMENTAL MATERIALS BY LASER SAMPLING INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
DENOYER, ER .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (08) :1187-1193
[10]   SEMIQUANTITATIVE SURVEY CAPABILITIES OF INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
EKIMOFF, D ;
VANNORSTRAND, AM ;
MOWERS, DA .
APPLIED SPECTROSCOPY, 1989, 43 (07) :1252-1257