Flow injection analysis for residual chlorine using Pb(II) ion-selective electrode detector

被引:22
作者
Sakai, A
Hemmi, A
Hachiya, H
Kobayashi, F
Ito, S
Asano, Y
Imato, T [1 ]
Fushinuki, Y
Taniguchi, I
机构
[1] Kyushu Univ, Grad Sch Engn, Dept Chem Syst Engn, Higashi Ku, Fukuoka 812, Japan
[2] DDK Corp, Div Res & Dev, Musashino, Tokyo 180, Japan
[3] Kagoshima Pref Police HQ, Sci Invest Res Lab, Kagoshima 892, Japan
[4] Kumamoto Univ, Kumamoto 860, Japan
关键词
flow injection analysis; residual chlorine; Pb(II) ion-selective electrode detector; tap water;
D O I
10.1016/S0039-9140(97)00299-3
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple flow injection analysis (FIA) system for residual chlorine in tap water has been developed by using a Pb(II) ion-selective electrode (ISE) detector. The method is based on a specific response of the Pb(II)-ISE to residual chlorine. The FIA system consists of a millivolt meter, a peristaltic pump, a Pb(II)-ISE detector and a recorder. A linear working curve between peak height and concentration of residual chlorine was obtained from 0.1 to 1 mg l(-1) for the developed FIA system. The relative standard deviation for repeated injections of a 0.2 mg l(-1) residual chlorine sample was 2%. The regression line and its correlation factor between the conventional o-tolidine colorimetric method and the present method were Y = 0.75X + 0.17 and 0.967, respectively, for this determination. (C) 1998 Elsevier Science B.V.
引用
收藏
页码:575 / 581
页数:7
相关论文
共 16 条
[1]  
*APHA AWWA WCCF, 1985, STAND METH EX WAT WA
[2]  
ASANO Y, 1979, ANN M JPN SOC AN CHE, V28, P729
[3]  
BEACH JS, 1977, PROG WATER TECH, V9, P97
[4]  
BUTTERFIELD CT, 1948, PUBLIC HLTH REPORTS, V3, P934
[5]  
CHARLOT G, 1957, ANAL QUALITATIVE R 2, pCH12
[6]   SEQUENTIAL INJECTION-ANALYSIS FOR ELECTROCHEMICAL MEASUREMENTS AND PROCESS ANALYSIS [J].
CHRISTIAN, GD .
ANALYST, 1994, 119 (11) :2309-2314
[7]  
CHRISTIAN GD, 1993, COMP REND 8 COLL PON, P7
[8]  
*DKK CORP, 1987, INSTR MAN PB II ION, P24
[9]   Improvement of the silver/silver chloride reference electrode and its application to pH measurement [J].
Ito, S ;
Hachiya, H ;
Baba, K ;
Asano, Y ;
Wada, H .
TALANTA, 1995, 42 (11) :1685-1690
[10]  
*JAP STAND ASS, 1994, 0101 JIS K, P50