Lung fluid and impedance cardiography

被引:9
作者
Critchley, LAH [1 ]
Critchley, JAJH
机构
[1] Chinese Univ Hong Kong, Prince Wales Hosp, Dept Anaesthesia & Intens Care, Hong Kong, Peoples R China
[2] Chinese Univ Hong Kong, Prince Wales Hosp, Dept Clin Pharmacol, Hong Kong, Peoples R China
关键词
complications; pulmonary oedema; measurement technique; impedance cardiography; thermodilution; monitoring; cardiac output;
D O I
10.1046/j.1365-2044.1998.00297.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Serial thermodilution and impedance cardiac output measurements in two critically ill patients are presented. Impedance cardiography failed to detect changes in cardiac output and provided values lower than those derived from thermodilution. Chest X-ray examination and changes in thoracic impedance suggested that these two patients had significantly increased extravascular lung water. This failure of impedance cardiography is attributed to aberrant electrical conduction though the lungs as a result of increased lung fluid that alters the impedance waveform. Although reliable when used in normal subjects, impedance cardiography appears not to provide accurate measurements in critically ill patients.
引用
收藏
页码:369 / 372
页数:4
相关论文
共 10 条
[1]   A NEW STROKE VOLUME EQUATION FOR THORACIC ELECTRICAL BIOIMPEDANCE - THEORY AND RATIONALE [J].
BERNSTEIN, DP .
CRITICAL CARE MEDICINE, 1986, 14 (10) :904-909
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Abdominal surgery alters the calibration of bioimpedance cardiac output measurement [J].
Critchley, LAH ;
Leung, DHY ;
Short, TG .
INTERNATIONAL JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1996, 13 (01) :1-8
[4]  
FULLER HD, 1992, CLIN INVEST MED, V15, P103
[5]   EVALUATION OF THE PORTABLE CHEST ROENTGENOGRAM FOR QUANTITATING EXTRAVASCULAR LUNG WATER IN CRITICALLY ILL ADULTS [J].
HALPERIN, BD ;
FEELEY, TW ;
MIHM, FG ;
CHILES, C ;
GUTHANER, DF ;
BLANK, NE .
CHEST, 1985, 88 (05) :649-652
[6]  
KUBICEK WG, 1966, AEROSPACE MED, V37, P1208
[7]   MULTICOMPONENT NONINVASIVE PHYSIOLOGIC MONITORING OF CIRCULATORY FUNCTION [J].
SHOEMAKER, WC ;
APPEL, PL ;
KRAM, HB ;
NATHAN, RC ;
THOMPSON, JL .
CRITICAL CARE MEDICINE, 1988, 16 (05) :482-490
[8]  
SHOEMAKER WC, 1994, CRIT CARE MED, V22, P1907
[9]   NONINVASIVE CARDIAC-OUTPUT - SIMULTANEOUS COMPARISON OF 2 DIFFERENT METHODS WITH THERMODILUTION [J].
WONG, DH ;
TREMPER, KK ;
STEMMER, EA ;
OCONNOR, D ;
WILBUR, S ;
ZACCARI, J ;
REEVES, C ;
WEIDOFF, P ;
TRUJILLO, RJ .
ANESTHESIOLOGY, 1990, 72 (05) :784-792
[10]   COMPARISON OF THORACIC ELECTRICAL BIOIMPEDANCE AND THERMODILUTION FOR THE MEASUREMENT OF CARDIAC INDEX IN PATIENTS WITH SEVERE SEPSIS [J].
YOUNG, JD ;
MCQUILLAN, P .
BRITISH JOURNAL OF ANAESTHESIA, 1993, 70 (01) :58-62