Plasma radiometry with 30 chord resolution for fast transients in the DIII-D tokamak

被引:14
作者
Gray, DS [1 ]
Hollmann, EM
Luckhardt, SC
Chalfant, J
Chousal, L
Hernandez, R
Jones, E
Kellman, AG
机构
[1] Univ Calif San Diego, Energy Res Ctr, La Jolla, CA 92093 USA
[2] Gen Atom Co, San Diego, CA 92186 USA
关键词
D O I
10.1063/1.1787149
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A diagnostic capable of providing time resolved measurements of plasma radiated power during disruptions and other fast transients, e.g., edge localized modes has been employed in the DIII-D tokamak. The radiation is detected with absolute extreme ultraviolet (AXUV) photodiode arrays. Thirty chords from a single port provide measurements from a full slice of the plasma at one toroidal location. The analog bandwidth is up to 1 MHz for the brightest events, i.e., disruptions. Active cooling of the diode arrays prevents damage during high temperature vessel baking. Effective responsivity values of 0.12-0.18 A/W are taken from previous work on the application of AXUV diodes in DIII-D. The total radiated energy in disruptions typically agrees with bolometer measurements within about 12%. (C) 2004 American Institute of Physics.
引用
收藏
页码:4133 / 4135
页数:3
相关论文
共 9 条
[1]   High resolution bolometry on the Alcator C-Mod tokamak (invited) [J].
Boivin, RL ;
Goetz, JA ;
Marmar, ES ;
Rice, JE ;
Terry, JL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (01) :260-264
[2]   Time resolved radiated power during tokamak disruptions and spectral averaging of AXUV photodiode response in DIII-D [J].
Gray, DS ;
Luckhardt, SC ;
Chousal, L ;
Gunner, G ;
Kellman, AG ;
Whyte, DG .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (02) :376-381
[3]   Present status of radiometric quality silicon photodiodes [J].
Korde, R ;
Prince, C ;
Cunningham, D ;
Vest, RE ;
Gullikson, E .
METROLOGIA, 2003, 40 (01) :S145-S149
[4]   THE EFFECT OF NEUTRON-IRRADIATION ON SILICON PHOTODIODES [J].
KORDE, R ;
OJHA, A ;
BRAASCH, R ;
ENGLISH, TC .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (06) :2169-2175
[5]   ONE GIGARAD PASSIVATING NITRIDED OXIDES FOR 100-PERCENT INTERNAL QUANTUM EFFICIENCY SILICON PHOTODIODES [J].
KORDE, R ;
CABLE, JS ;
CANFIELD, LR .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (06) :1655-1659
[6]   2D TOMOGRAPHY WITH BOLOMETRY IN DIII-D [J].
LEONARD, AW ;
MEYER, WH ;
GEER, B ;
BEHNE, DM ;
HILL, DN .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (02) :1201-1204
[7]   WIDE-BAND SILICON BOLOMETERS ON THE LSX FIELD REVERSED CONFIGURATION EXPERIMENT [J].
MAQUEDA, RJ ;
WURDEN, GA ;
CRAWFORD, EA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (10) :4717-4719
[8]   Multilayer mirror and foil filter AXUV diode arrays on CDX-U spherical torus [J].
Soukhanovskii, VA ;
Stutman, D ;
Iovea, M ;
Finkenthal, M ;
Moos, HW ;
Munsat, T ;
Jones, B ;
Hoffman, D ;
Kaita, R ;
Majeski, R .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (01) :737-741
[9]   HIGH-SENSITIVITY, HIGH-RESOLUTION MEASUREMENTS OF RADIATED POWER ON TEXT-U [J].
WEN, YZ ;
BRAVENEC, RV .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (01) :549-551