Continuous differential monitoring of the spent dialysate glucose level: clinical evaluation

被引:5
作者
Deman, P [1 ]
Suls, J [1 ]
Sansen, W [1 ]
机构
[1] Katholieke Univ Leuven, ESAT, MICAS, B-3001 Louvain, Belgium
关键词
biosensors; enzymatic; glucose; interferents; potentiostat;
D O I
10.1016/S0925-4005(97)00222-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new glucose monitoring system for the haemodialysis is presented. This new tool measures real-time glucose concentrations in the dialysate flow of the artificial kidney. The background signal coming from present clinical interferents is effectively rejected by using a differential measurement set-up. The amperometric device is based on thick film RuO2 sensors. They are connected ta a novel portable system with an optimal patient safety protection. This is guaranteed by means of a galvanic-isolated power supply and an optically-isolated data-communication to a lap-top PC. The sensor interface can easily be controlled via a graphical user interface-screen. (C) 1997 Elsevier Science S.A.
引用
收藏
页码:304 / 308
页数:5
相关论文
共 22 条
  • [1] DEMAN P, 1996, THESIS KU LEUVEN
  • [2] DEMAN P, 1995, THESIS PROVINCIALE I
  • [3] DEVEY GB, 1974, MED ENG YB MED 4, P1090
  • [4] ELECTRICITY IN HOSPITALS - ELIMINATION OF LETHAL HAZARDS
    FRIEDLANDER, GD
    [J]. IEEE SPECTRUM, 1971, 8 (09) : 40 - +
  • [5] HANNING I, 1988, ANAL LETT, V19, P461
  • [6] HNATEK ER, 1972, DESIGN SOLID STATE P
  • [7] PERFORMANCE OF A PLANAR DIFFERENTIAL-CONDUCTIVITY SENSOR FOR UREA
    JACOBS, P
    SULS, J
    SANSEN, W
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1994, 20 (2-3) : 193 - 198
  • [8] Jacobs P, 1993, ASAIO J, V39, pM353
  • [9] Jacobs P, 1994, ASAIO J, V40, pM393, DOI 10.1097/00002480-199407000-00029
  • [10] Lambrechts, 1992, BIOSENSORS MICROELEC