Detectors and detection limits in INAA - I. General theoretical relationships between detector specifications and detection limits

被引:3
作者
Bode, P [1 ]
机构
[1] Delft Univ Technol, Interfac Reactor Inst, NL-2629 JB Delft, Netherlands
关键词
D O I
10.1007/BF02034257
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Expressions have been derived which link detection limits in INAA with the Ge-detector specifications, provided by the vendors: relative efficiency, peak-to-Compton ratio and resolution. Situations have been distinguished in which a peak has to be detected on a Compton background, or on a natural background. The expressions allow for a direct indication of the improvement in detection limits, or the shortening of turnaround-time to attain equal detection limits, when a given detector is replaced by a detector with better specifications.
引用
收藏
页码:117 / 125
页数:9
相关论文
共 17 条
[1]   PERFORMANCES OF LARGE VOLUME P-TYPE HPGE DETECTORS [J].
BELLIA, G ;
DELZOPPO, A ;
AGODI, C ;
ALBA, R ;
CONIGLIONE, R ;
MIGNECO, E ;
RUSSO, G ;
SAPIENZA, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 284 (2-3) :399-404
[2]  
BLAAUW M, 1993, THESIS DELFT U TECHN
[3]  
Bode P., 1989, Transactions of the American Nuclear Society, V60, P6
[4]   ADVANCED DETECTOR SYSTEMS - WHAT DO THEY HAVE TO OFFER FOR ACTIVATION-ANALYSIS [J].
BODE, P ;
LINDSTROM, RM .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1993, 167 (01) :187-195
[5]  
BODE P, IN PRESS
[6]  
BRAUER FP, 1978, IEEE TRANSACT, P398
[7]   FACTORS DETERMINING ULTIMATE DETECTION SENSITIVITY OF GE(LI) GAMMA-RAY SPECTROMETERS [J].
COOPER, JA .
NUCLEAR INSTRUMENTS & METHODS, 1970, 82 :273-+
[9]   APPLICATION OF ANTI-COMPTON COUNTING IN INSTRUMENTAL NEUTRON-ACTIVATION ANALYSIS [J].
DAS, HA ;
ZONDERHUIS, J .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1987, 114 (02) :207-213
[10]  
*EG ORTEC, 1995, 1993 94 CAT INSTR SY, P475