Comparison of eight commercial kits for quantitation of antineutrophil cytoplasmic antibodies (ANCA)

被引:32
作者
Wang, GC
Csernok, E [2 ]
de Groot, K
Gross, WL
机构
[1] Rheumaklin Bad Bramstedt GmbH, D-24576 Bad Bramstedt, Germany
[2] Univ Lubeck, Dept Rheumatol, D-2400 Lubeck, Germany
关键词
ANCA; PR3-ANCA; MPO-ANCA; ELISA; indirect immunofluorescence;
D O I
10.1016/S0022-1759(97)00154-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Antineutrophil cytoplasmic antibodies (ANCA) are used as diagnostic markers for systemic vasculitis. However,the specificity and sensitivity of ANCA detection differs from centre to centre due in large part to variations in methodology. We compared 8 commercial ELISA kits and an in-house method (HM) for their specificity and sensitivity in detecting ANCA against proteinase 3 (PR3-ANCA, 7 kits) and meyloperoxidse (MPO-ANCA. 8 kits). Sera from 5 patients with systemic lupus erythematosus (SLE), 28 with Wegener's granulomatosis (WG), 22 with microscopic polyangiitis (MPA), 5 with idiopathic rapidly progressive glomerulonephritis (RPGN), and 5 healthy controls were examined by both the indirect immunofluorescence technique (IFT) and the ELISA kits. Sera from healthy controls and patients with SLE or cANCA-negative WG were shown to be PR3-ANCA negative by all 7 PR3-ANCA kits. In 25 cANCA-positive sera from WG patients, PR3-ANCA positivity ranged from 44% to 84%. An absolute concordance among the 7 kits was noted in 56% of the cANCA-positive samples. The PR3-ANCA levels in 5 of the 7 kits correlated with the cANCA titers in IFT. Sera from the healthy controls and 4 out of the 5 SLE and pANCA-negative patients were found to be MPO-ANCA negative in all 8 MPO-ANCA kits. In 20 pANCA-positive sera, MPO-ANCA positivity ranged from 25% to 75%. Thirty-five percent of MPO-ANCA-positive sera were confirmed by capture ELISA, immunoblot and inhibition assay. The concordance rate was only 30% among pANCA-positive sera in the 8 MPO-ANCA kits. No significant correlation was observed between pANCA titers and MPO-ANCA levels. The HM showed that 65% of cANCA-positive sera were PR3-ANCA positive, and 45% of pANCA-positive sera were MPO-ANCA positive. Our results indicate that the sensitivities and specificities for ANCA detection differ significantly among the commercial kits tested and underline the necessity of establishing uniform international standards for ANCA ELISA procedures in order to permit more reliable interpretation and comparison of data. (C) 1997 Elsevier Science B.V.
引用
收藏
页码:203 / 211
页数:9
相关论文
共 16 条
[1]   ANTILACTOFERRIN ANTIBODIES IN PATIENTS WITH RHEUMATOID-ARTHRITIS ARE ASSOCIATED WITH VASCULITIS [J].
COREMANS, IEM ;
HAGEN, EC ;
DAHA, MR ;
VANDERWOUDE, FJ ;
VANDERVOORT, EAM ;
KLEIJBURGVANDERKEUR, C ;
BREEDVELD, FC .
ARTHRITIS AND RHEUMATISM, 1992, 35 (12) :1466-1475
[2]  
DAVENPORT A, 1994, Q J MED, V87, P291
[3]   ANTI-NEUTROPHIL CYTOPLASMIC AUTOANTIBODIES WITH SPECIFICITY FOR MYELOPEROXIDASE IN PATIENTS WITH SYSTEMIC VASCULITIS AND IDIOPATHIC NECROTIZING AND CRESCENTIC GLOMERULONEPHRITIS [J].
FALK, RJ ;
JENNETTE, JC .
NEW ENGLAND JOURNAL OF MEDICINE, 1988, 318 (25) :1651-1657
[4]  
Gross Wolfgang L., 1995, Current Opinion in Rheumatology, V7, P11
[5]  
HAGEN EC, 1993, J IMMUNOL METHODS, V159, P1
[6]  
HAGEN EC, IN PRESS NEW ENGL J
[7]   NOMENCLATURE OF SYSTEMIC VASCULITIDES - PROPOSAL OF AN INTERNATIONAL CONSENSUS CONFERENCE [J].
JENNETTE, JC ;
FALK, RJ ;
ANDRASSY, K ;
BACON, PA ;
CHURG, J ;
GROSS, WL ;
HAGEN, EC ;
HOFFMAN, GS ;
HUNDER, GG ;
KALLENBERG, CGM ;
MCCLUSKEY, RT ;
SINICO, RA ;
REES, AJ ;
VANES, LA ;
WALDHERR, R ;
WIIK, A .
ARTHRITIS AND RHEUMATISM, 1994, 37 (02) :187-192
[8]  
KALLENBERG CGM, 1996, VASCULITIDES, P48
[9]   THE AMERICAN-COLLEGE-OF-RHEUMATOLOGY 1990 CRITERIA FOR THE CLASSIFICATION OF WEGENER GRANULOMATOSIS [J].
LEAVITT, RY ;
FAUCI, AS ;
BLOCH, DA ;
MICHEL, BA ;
HUNDER, GG ;
AREND, WP ;
CALABRESE, LH ;
FRIES, JF ;
LIE, JT ;
LIGHTFOOT, RW ;
MASI, AT ;
MCSHANE, DJ ;
MILLS, JA ;
STEVENS, MB ;
WALLACE, SL ;
ZVAIFLER, NJ .
ARTHRITIS AND RHEUMATISM, 1990, 33 (08) :1101-1107
[10]   ANTICYTOPLASMIC AUTOANTIBODIES - THEIR IMMUNODIAGNOSTIC VALUE IN WEGENER GRANULOMATOSIS [J].
NOLLE, B ;
SPECKS, U ;
LUDEMANN, J ;
ROHRBACH, MS ;
DEREMEE, RA ;
GROSS, WL .
ANNALS OF INTERNAL MEDICINE, 1989, 111 (01) :28-40