Effects of gradient anisotropy in MRI

被引:46
作者
Aldefeld, B [1 ]
Börnert, P [1 ]
机构
[1] Philips Res, Div Tech Syst, D-22335 Hamburg, Germany
关键词
MRI; gradients; eddy currents; artifacts;
D O I
10.1002/mrm.1910390414
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A gradient system is anisotropic if the impulse responses of at least two of the gradient channels, x, y, or z, differ from each other. Such an undesired condition may arise, for example, from differences between the gradient channels with respect to eddy currents or from unbalanced time delays in the electronic components. Depending on the degree of anisotropy, the actual gradient then deviates from the nominal, desired gradient under certain oblique orientations during the transient periods of gradient switching, The adverse consequence is degradation of image quality, such as distortion, ghosting, and blurring. In this paper, a theoretical analysis is given of the basic effects. Furthermore, the implications for the MRI process and possible correction methods are described. The effects of anisotropy are shown experimentally for echo-planar imaging and two-dimensional selective RF excitation with spiral gradient pulses.
引用
收藏
页码:606 / 614
页数:9
相关论文
共 11 条
[1]   HIGH-SPEED SPIRAL-SCAN ECHO PLANAR NMR IMAGING .1. [J].
AHN, CB ;
KIM, JH ;
CHO, ZH .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1986, 5 (01) :2-7
[2]  
ALDEFELD B, 1997, P ISMRM 5 SCI M VANC, P1835
[3]  
BORNERT P, 1997, P ISMRM 5 SCI M VANC, P1534
[4]   IMAGE-RECONSTRUCTION FOR ECHO PLANAR IMAGING WITH NONEQUIDISTANT KAPPA-SPACE SAMPLING [J].
BRUDER, H ;
FISCHER, H ;
REINFELDER, HE ;
SCHMITT, F .
MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (02) :311-323
[5]  
GATEHOUSE PD, 1996, P ISMRM 4 ANN M NEW, P1481
[6]   RARE IMAGING - A FAST IMAGING METHOD FOR CLINICAL MR [J].
HENNIG, J ;
NAUERTH, A ;
FRIEDBURG, H .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (06) :823-833
[7]   A K-SPACE ANALYSIS OF SMALL-TIP-ANGLE EXCITATION [J].
PAULY, J ;
NISHIMURA, D ;
MACOVSKI, A .
JOURNAL OF MAGNETIC RESONANCE, 1989, 81 (01) :43-56
[8]  
PUALY J, 1989, P SMRM 8 ANN M AMST, P28
[9]   COMPENSATION OF MULTIDIMENSIONAL SELECTIVE EXCITATION PULSES USING MEASURED K-SPACE TRAJECTORIES [J].
TAKAHASHI, A ;
PETERS, T .
MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (03) :446-456
[10]  
ZHOU X, 1996, P ISMRM 4 SCI M NEW, P1477