Development of total uranium analytical method by L X-ray fluorescence

被引:5
作者
Dewberry, RA [1 ]
机构
[1] Westinghouse Savannah River Co, Savannah River Technol Ctr, Aiken, SC 29801 USA
关键词
X-ray fluorescence; U L X-rays; self-absorption correction;
D O I
10.1016/S0168-9002(97)01113-3
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This paper describes development of an L X-ray fluorescence technique to perform total uranium analysis using an internal excitation source which is added directly to the sample. The method has been demonstrated with synthetic U samples in the limited concentration range 1-15 g/l, and provides the advantage of simplicity, involving no mechanical parts which would normally be found in an external excitation source. Total uranium is determined by counting L X-rays fluoresced by a microCurie level spike of Cd-109 added directly to the sample and without shielding the excitation source from the detector. A method for correction of sample self-absorption is included in the analysis. (C) 1998 Published by Elsevier Science B.V. All rights reserved.