Safety concerns in ultrahigh pressure capillary liquid chromatography using air-driven pumps

被引:32
作者
Xiang, YQ
Maynes, DR
Lee, ML [1 ]
机构
[1] Brigham Young Univ, Dept Chem & Biochem, Provo, UT 84602 USA
[2] Brigham Young Univ, Dept Mech Engn, Provo, UT 84602 USA
关键词
ultrahigh-pressure liquid chromatography; air-driven pumps; safety concerns;
D O I
10.1016/S0021-9673(03)00171-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Ultrahigh pressure liquid chromatography (UHPLC) is an emerging technique which utilizes pressures higher than 10 000 p.s.i. to overcome the flow resistance imposed when using very small particles as packing materials in fused-silica capillary columns (1 p.s.i. = 6894.76 Pa). This technique has demonstrated exceptionally high separation speeds and chromatographic efficiencies. However, safety is a concern when extremely high pressures are used. In this study, the safety aspects of capillary column rupture during operation were identified and carefully evaluated. First, liquid jets may be formed as a result of blow-out of the on-column frits or from rupture of the capillary at or near the column inlet. Second, incorrect installation of the capillary at the injector. failure of the ferrule used in the capillary connection, or rupture of the capillary can produce high speed projectiles of silica particles or column fragments. Experiments were carried out in the laboratory to produce liquid (water) jets and capillary projectiles using a UHPLC system, and the power density, an important parameter describing water jets in industrial practice, was calculated. Experimental results were in accordance with theoretical calculations. Both indicated that water jets and capillary projectiles under ultrahigh pressures might lead to skin penetration under limited conditions. The use of a plexiglass shroud to cover an initial length of the installed capillary column can eliminate any safety-related concerns about liquid jets or capillary projectiles. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:189 / 196
页数:8
相关论文
共 20 条
[1]  
Axmann HD, 1996, BHR GR CONF SER PUBL, P309
[2]  
AXMANN HD, 1994, P 12 INT C JETT TECH, P317
[3]   TREATMENT OF HIGH-PRESSURE WATER GUN INJECTION INJURY OF THE FOOT WITH ADJUNCTIVE HYPERBARIC-OXYGEN - A CASE-REPORT [J].
CALHOUN, JH ;
GOGAN, WJ ;
VIEGAS, SF ;
MADER, JT .
FOOT & ANKLE, 1989, 10 (01) :40-42
[4]   NEWTONIAN JET STABILITY [J].
GRANT, RP ;
MIDDLEMA.S .
AICHE JOURNAL, 1966, 12 (04) :669-&
[5]  
Honl M, 2000, BHR GR CONF SER PUBL, P183
[6]   Practical examination of a nonporous silica stationary phase for reversed-phase fast LC applications [J].
Jenke, DR .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1996, 34 (08) :362-367
[7]  
Lippert JA, 1999, J MICROCOLUMN SEP, V11, P631, DOI 10.1002/(SICI)1520-667X(199911)11:9<631::AID-MCS1>3.3.CO
[8]  
2-9
[9]   Ultrahigh pressure reversed-phase capillary liquid chromatography:: Isocratic and gradient elution using columns packed with 1.0-μm particles [J].
MacNair, JE ;
Patel, KD ;
Jorgenson, JW .
ANALYTICAL CHEMISTRY, 1999, 71 (03) :700-708
[10]   Ultrahigh pressure reversed-phase liquid chromatography in packed capillary columns [J].
MacNair, JE ;
Lewis, KC ;
Jorgenson, JW .
ANALYTICAL CHEMISTRY, 1997, 69 (06) :983-989