Residual gentamicin-release from antibiotic-loaded polymethylmethacrylate beads after 5 years of implantation

被引:153
作者
Neut, D
van de Belt, H
van Horn, JR
van der Mei, HC
Busscher, HJ
机构
[1] Univ Groningen, Dept Biomed Engn, NL-9713 AV Groningen, Netherlands
[2] Univ Groningen Hosp, Dept Orthoped Surg, NL-9713 GZ Groningen, Netherlands
[3] Streekziekenhuis Coevorden Hardenberg, Dept Orthoped Surg, NL-7772 SE Hardenberg, Netherlands
关键词
polymethylmethacryate; beads; gentamicin; biodegradable; antibiotic resistance;
D O I
10.1016/S0142-9612(02)00614-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In infected joint arthroplasty, high local levels of antibiotics are achieved through temporary implantation of non-biodegradable gentamicin-loaded polymethylmethacrylate beads. Despite their antibiotic release, these beads act as a biomaterial surface to which bacteria preferentially adhere, grow and potentially develop antibiotic resistance. In routine clinical practice, these beads are removed after 14 days, but for a variety of reasons, we were confronted with a patient in which these beads were left in situ for 5 years. Retrieval of gentamicin-loaded beads from this patient constituted an exceptional case to study the effects of long-term implantation on potentially colonizing microflora and gentamicin release. Gentamicin-release test revealed residual antibiotic release after being 5 years in situ and extensive microbiological sampling resulted in recovery of a gentamicin-resistant staphylococcal strain from the bead surface. This case emphasizes the importance of developing biodegradable antibiotic-loaded beads Lis an antibiotic delivery system. (C) 2003 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:1829 / 1831
页数:3
相关论文
共 11 条
[1]   In vitro elution of tobramycin from bioabsorbable polycaprolactone beads [J].
Burd, TA ;
Anglen, JO ;
Lowry, KJ ;
Hendricks, KJ ;
Day, D .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2001, 15 (06) :424-428
[2]   BIOMATERIAL-CENTERED INFECTION - MICROBIAL ADHESION VERSUS TISSUE INTEGRATION [J].
GRISTINA, AG .
SCIENCE, 1987, 237 (4822) :1588-1595
[3]  
KLEMM KW, 1993, CLIN ORTHOP RELAT R, P63
[4]  
NELSON CL, 1994, ORTHOPEDICS, V17, P415
[5]   Biomaterial-associated infection of gentamicin-loaded PMMA beads in orthopaedic revision surgery [J].
Neut, D ;
van de Belt, H ;
Stokroos, I ;
van Horn, JR ;
van der Mei, HC ;
Busscher, HJ .
JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 2001, 47 (06) :885-891
[6]   ... or not to treat? [J].
van de Belt, H ;
Neut, D ;
van Horn, JR ;
van der Mei, HC ;
Schenk, W ;
Busscher, HJ .
NATURE MEDICINE, 1999, 5 (04) :358-359
[7]   Isolation of small colony variants of Staphylococcus aureus following gentamicin bead placement for osteomyelitis as a possible cause for recurrence [J].
von Eiff, C ;
Lindner, N ;
Proctor, RA ;
Winkemann, W ;
Peters, G .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1998, 136 (03) :268-271
[8]  
WAHLIG H, 1979, ZBL CHIR, V104, P923
[9]  
WALENKAMP GHIM, 1986, CLIN ORTHOP RELAT R, P171
[10]   How I do it - Chronic osteomyelitis [J].
Walenkamp, GHIM .