Room temperature sonolysis-based advanced oxidation process for degradation of organomercurials: Application to determination of inorganic and total mercury in waters by flow injection-cold vapor atomic absorption spectrometry

被引:73
作者
Capelo, JL [1 ]
Lavilla, I [1 ]
Bendicho, C [1 ]
机构
[1] Univ Vigo, Dept Quim Analit & Alimentaria, Fac Ciencias Quim, Vigo 36200, Spain
关键词
D O I
10.1021/ac000470b
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new oxidation method based on room-temperature ultrasonic irradiation (sonolysis) is proposed for Conversion of organomercurials into inorganic mercury and subsequent determination by now injection-cold vapor atomic absorption spectrometry. This advanced oxidation process eliminates the need for chemical oxidants, high temperature, and pressure for degradation of organomercurials so that total mercury can be determined with sodium tetrahydroborate(III) or tin(II) chloride as reducing agents. Complete oxidations can be accomplished within 3 min, using a 40% sonication amplitude (100 W nominal power) provided by a probe ultrasonic device (20 kHz frequency) and a 1. mol L-1 HCl liquid medium. The presence of HCl was seen to be necessary for fast oxidation of organomercurials, in contrast to other chemical oxidants such as H2O2 or HNO3 which yielded incomplete oxidation. Further advantages of the proposed method over existing methods which are currently employed for oxidation prior to total Hg determination are the removal of hazardous wastes and the decreased risk of Hg losses by volatilization. Oxidation kinetics indicated a pseudo first-order reaction with apparent rate constants (k) of 3.2 x 10(-2) and 1.6 x 10(-2) s(-1) for methylmercury and phenylmercury, respectively. Oxidation experiments in the presence of foreign substances acting as OH radical scavengers showed a tolerance at least up to a concentration of 1000 mg L-1. likewise, model wastewaters with chemical oxygen demand of up to 1000 mg L-1 could be processed without diminishing the oxidation efficiency. The method was applied to determination of inorganic and total mercury in simulated wastewaters and spiked environmental waters in combination with selective reduction.
引用
收藏
页码:4979 / 4984
页数:6
相关论文
共 42 条
[1]   DECOMPOSITION AND STABILITY STUDIES OF METHYLMERCURY IN WATER USING COLD VAPOR ATOMIC-ABSORPTION SPECTROMETRY [J].
AHMED, R ;
STOEPPLER, M .
ANALYST, 1986, 111 (12) :1371-1374
[2]   CRITICAL COMPARISON OF 2 STANDARD DIGESTION PROCEDURES FOR THE DETERMINATION OF TOTAL MERCURY IN NATURAL-WATER SAMPLES BY COLD VAPOR ATOMIC-ABSORPTION SPECTROMETRY [J].
BAXTER, DC ;
FRECH, W .
ANALYTICA CHIMICA ACTA, 1990, 236 (02) :377-384
[3]   AUTOMATED CONTINUOUS MONITORING OF INORGANIC AND TOTAL MERCURY IN WASTE-WATER AND OTHER WATERS BY FLOW-INJECTION ANALYSIS AND COLD-VAPOR ATOMIC-ABSORPTION SPECTROMETRY [J].
BIRNIE, SE .
JOURNAL OF AUTOMATIC CHEMISTRY, 1988, 10 (03) :140-143
[4]   Determination of mercury in filtered sea-water by flow injection with on-line oxidation and atomic fluorescence spectrometric detection [J].
Bloxham, MJ ;
Hill, SJ ;
Worsfold, PJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (07) :511-514
[5]   SONOCHEMICAL DESTRUCTION OF CHLORINATED HYDROCARBONS IN DILUTE AQUEOUS-SOLUTION [J].
CHEUNG, HM ;
BHATNAGAR, A ;
JANSEN, G .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1991, 25 (08) :1510-1512
[6]   ULTRASONIC NEBULIZATION AND ARSENIC VALENCE STATE CONSIDERATIONS PRIOR TO DETERMINATION VIA INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
CREED, JT ;
MARTIN, TD ;
BROCKHOFF, CA .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1995, 10 (06) :443-447
[7]   Determination of mercury in saliva with a flow-injection system [J].
Guo, T ;
Baasner, J ;
Gradl, M ;
Kistner, A .
ANALYTICA CHIMICA ACTA, 1996, 320 (2-3) :171-176
[8]   DETERMINATION OF MERCURY IN URINE BY FLOW-INJECTION COLD VAPOR ATOMIC-ABSORPTION SPECTROMETRY [J].
GUO, TZ ;
BAASNER, J .
ANALYTICA CHIMICA ACTA, 1993, 278 (01) :189-196
[9]   Technical note: Determination of mercury in blood by on-line digestion with FIMS [J].
Guo, TZ ;
Baasner, J .
JOURNAL OF AUTOMATIC CHEMISTRY, 1996, 18 (06) :217-220
[10]   ONLINE MICROWAVE SAMPLE PRETREATMENT FOR THE DETERMINATION OF MERCURY IN BLOOD BY FLOW-INJECTION COLD VAPOR ATOMIC-ABSORPTION SPECTROMETRY [J].
GUO, TZ ;
BAASNER, J .
TALANTA, 1993, 40 (12) :1927-1936