Influence of layered tissue architecture on estimates of tissue optical properties obtained from spatially resolved diffuse reflectometry

被引:133
作者
Farrell, TJ
Patterson, MS
Essenpreis, M
机构
[1] Boehringer Mannheim GmbH, D-68305 Mannheim, Germany
[2] Hamilton Reg Canc Ctr, Hamilton, ON L8V 5C2, Canada
[3] McMaster Univ, Hamilton, ON L8V 5C2, Canada
来源
APPLIED OPTICS | 1998年 / 37卷 / 10期
关键词
D O I
10.1364/AO.37.001958
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Most instruments used to measure tissue optical properties noninvasively employ data-analysis algorithms that rely on the simplifying assumption that the tissue is semi-infinite and homogeneous. The influence of a layered tissue architecture on the determination of the scattering and absorption coefficients has been investigated in this study. Reflectance as a function of distance from a point source for a two-layered tissue architecture that simulates skin overlying fat was calculated by using a Monte Carlo code. These data were analyzed by using a diffusion theory model for a homogeneous semi-infinite medium to calculate the scatter and absorption coefficients. Depending on the algorithm and the radial distance, the estimated tissue optical properties were different from those of either layer, and under some circumstances, physically impossible. In addition, the sensitivity and cross talk of the estimated optical properties to changes in input optical properties were calculated for different layered geometries. For typical optical properties of skin, the sensitivity to changes in optical properties is highly dependent on the layered architecture, the measurement distance, and the fitting algorithm. Furthermore, a change in the input absorption coefficient may result in an apparent change in the measured scatter coefficient, and a change in the input scatter coefficient may result in an apparent change in the measured absorption coefficient. (C) 1998 Optical Society of America.
引用
收藏
页码:1958 / 1972
页数:15
相关论文
共 29 条
[1]  
[Anonymous], 1990, APPL NUMERICAL ANAL
[2]  
Bevington R., 1969, DATA REDUCTION ERROR
[3]   Correlation between blood glucose concentration in diabetics and noninvasively measured tissue optical scattering coefficient [J].
Bruulsema, JT ;
Hayward, JE ;
Farrell, TJ ;
Patterson, MS ;
Heinemann, L ;
Berger, M ;
Koschinsky, T ;
SandahlChristiansen, J ;
Orskov, H .
OPTICS LETTERS, 1997, 22 (03) :190-192
[4]   SYSTEM FOR LONG-TERM MEASUREMENT OF CEREBRAL BLOOD AND TISSUE OXYGENATION ON NEWBORN-INFANTS BY NEAR-INFRARED TRANS-ILLUMINATION [J].
COPE, M ;
DELPY, DT .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1988, 26 (03) :289-294
[5]   PHOTON MIGRATION IN A 2-LAYER TURBID MEDIUM - A DIFFUSION ANALYSIS [J].
DAYAN, I ;
HAVLIN, S ;
WEISS, GH .
JOURNAL OF MODERN OPTICS, 1992, 39 (07) :1567-1582
[6]   QUANTITATIVE-DETERMINATION OF THE ABSORPTION-SPECTRA OF CHROMOPHORES IN STRONGLY SCATTERING MEDIA - A LIGHT-EMITTING-DIODE BASED TECHNIQUE [J].
FANTINI, S ;
FRANCESCHINI, MA ;
FISHKIN, JB ;
BARBIERI, B ;
GRATTON, E .
APPLIED OPTICS, 1994, 33 (22) :5204-5213
[7]   THE USE OF A NEURAL NETWORK TO DETERMINE TISSUE OPTICAL-PROPERTIES FROM SPATIALLY RESOLVED DIFFUSE REFLECTANCE MEASUREMENTS [J].
FARRELL, TJ ;
WILSON, BC ;
PATTERSON, MS .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (12) :2281-2286
[8]   A DIFFUSION-THEORY MODEL OF SPATIALLY RESOLVED, STEADY-STATE DIFFUSE REFLECTANCE FOR THE NONINVASIVE DETERMINATION OF TISSUE OPTICAL-PROPERTIES INVIVO [J].
FARRELL, TJ ;
PATTERSON, MS ;
WILSON, B .
MEDICAL PHYSICS, 1992, 19 (04) :879-888
[9]  
HASKELL RC, 1994, J OPT SOC AM A, V11, P2724
[10]   Diffuse radiation in the galaxy [J].
Henyey, LG ;
Greenstein, JL .
ASTROPHYSICAL JOURNAL, 1941, 93 (01) :70-83