Fast CCD camera for x-ray photon correlation spectroscopy and time-resolved x-ray scattering and imaging

被引:52
作者
Falus, P [1 ]
Borthwick, MA
Mochrie, SGJ
机构
[1] Yale Univ, Dept Phys, New Haven, CT 06520 USA
[2] MIT, Dept Phys, Cambridge, MA 02139 USA
[3] Yale Univ, Dept Appl Phys, New Haven, CT 06520 USA
关键词
D O I
10.1063/1.1808913
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A new, fast x-ray detector system is presented for high-throughput, high-sensitivity, time-resolved, x-ray scattering and imaging experiments, most especially x-ray photon correlation spectroscopy (XPCS). After a review of the architectures of different CCD chips and a critical examination of their suitability for use in a fast x-ray detector, the new detector hardware is described. In brief, its principal component is an inexpensive, commercial camera-the SMD1M60-originally designed for optical applications, and modified for use as a direct-illumination x-ray detector. The remainder of the system consists of two Coreco Imaging PC-DIG frame grabber boards, located inside a Dell Power-edge 6400 server. Each frame grabber sits on its own PCI bus and handles data from 2 of the CCD's 4 taps. The SMD1M60 is based on a fast, frame-transfer, 4-tap CCD chip, read out at12-bit resolution at frame rates of up to 62 Hz for full frame readout and up to 500 Hz for one-sixteenth frame readout. Experiments to characterize the camera's suitability for XPCS and small-angle x-ray scattering (SAXS) are presented. These experiments show that single photon events are readily identified, and localized to within a pixel index or so. This is a sufficiently fine spatial resolution to maintain the speckle contrast at an acceptable value for XPCS measurements. The detective quantum efficiency of the SMD1M60 is 49% for directly-detected 6.3 keV x rays. The effects of data acquisition strategies that permit near-real-time data compression are also determined and discussed. Overall, the SMD1M60 detector system represents a major improvement in the technology for time-resolved x-ray experiments, that require an area detector with time-resolutions in few-milliseconds-to-few-seconds range, and it should have wide applications, extending beyond XPCS. (C) 2004 American Institute of Physics.
引用
收藏
页码:4383 / 4400
页数:18
相关论文
共 35 条
[1]   EFFECTS OF SOFT-X-RAY IRRADIATION ON SOLID-STATE IMAGERS [J].
ALLINSON, NM ;
ALLSOPP, DWE ;
QUAYLE, JA ;
MAGORRIAN, BG .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 310 (1-2) :267-272
[2]   DEVELOPMENT OF NON-INTENSIFIED CHARGE-COUPLED-DEVICE AREA X-RAY-DETECTORS [J].
ALLINSON, NM .
JOURNAL OF SYNCHROTRON RADIATION, 1994, 1 :54-62
[3]   Pile-up on X-ray CCD instruments [J].
Ballet, J .
ASTRONOMY & ASTROPHYSICS SUPPLEMENT SERIES, 1999, 135 (02) :371-381
[4]   X-RAY-INTENSITY FLUCTUATION SPECTROSCOPY OBSERVATIONS OF CRITICAL-DYNAMICS IN FE3AL [J].
BRAUER, S ;
STEPHENSON, GB ;
SUTTON, M ;
BRUNING, R ;
DUFRESNE, E ;
MOCHRIE, SGJ ;
GRUBEL, G ;
ALSNIELSEN, J ;
ABERNATHY, DL .
PHYSICAL REVIEW LETTERS, 1995, 74 (11) :2010-2013
[5]   Ultralow-angle dynamic light scattering with a charge coupled device camera based multispeckle, multitau correlator [J].
Cipelletti, L ;
Weitz, DA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (08) :3214-3221
[6]   X-RAY PHOTON-CORRELATION SPECTROSCOPY STUDY OF BROWNIAN-MOTION OF GOLD COLLOIDS IN GLYCEROL [J].
DIERKER, SB ;
PINDAK, R ;
FLEMING, RM ;
ROBINSON, IK ;
BERMAN, L .
PHYSICAL REVIEW LETTERS, 1995, 75 (03) :449-452
[7]   A STATISTICAL TECHNIQUE FOR CHARACTERIZING X-RAY POSITION-SENSITIVE DETECTORS [J].
DUFRESNE, E ;
BRUNING, R ;
SUTTON, M ;
RODRICKS, B ;
STEPHENSON, GB .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 364 (02) :380-393
[8]   Concentration fluctuations in the binary mixture hexane-nitrobenzene with static and dynamic x-ray scattering [J].
Dufresne, EM ;
Nurushev, T ;
Clarke, R ;
Dierker, SB .
PHYSICAL REVIEW E, 2002, 65 (06) :1-061507
[9]  
FALUS P, IN PRESS J SYNCHROTR
[10]   Charge-coupled device area x-ray detectors [J].
Gruner, SM ;
Tate, MW ;
Eikenberry, EF .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (08) :2815-2842