Simultaneous determination of losartan, EXP-3174 and hydrochlorothiazide in plasma via fully automated 96-well-format-based solid-phase extraction and liquid chromatography-negative electrospray tandem mass spectrometry

被引:30
作者
Kolocouri, Filomila [1 ]
Dotsikas, Yannis [1 ]
Apostolou, Constantinos [1 ]
Kousoulos, Constantinos [1 ]
Loukas, Yannis L. [1 ]
机构
[1] Univ Athens, Lab Pharmaceut Anal & Bioequivalence Serv GLP Com, Dept Pharmaceut Chem, Sch Pharm, GR-15771 Athens, Greece
关键词
losartan; hydrochlorothiazide; solid-phase extraction; mass spectrometry; 96 deep well;
D O I
10.1007/s00216-006-0990-4
中图分类号
Q5 [生物化学];
学科分类号
071010 [生物化学与分子生物学]; 081704 [应用化学];
摘要
An automated, sensitive and high-throughput liquid chromatographic/electrospray tandem mass spectrometric (LC-MS/MS) assay was developed for the simultaneous determination of losartan (LOS), its major circulating metabolite EXP-3174 and hydrochlorothiazide (HCTZ) in human plasma. LOS and HCTZ coexist in the same drug formulation, and this is the first method that enables the simultaneous determination of both drugs along with the active metabolite of LOS. Since these drugs have different physicochemical properties, the employment of a liquid-liquid extraction (LLE) protocol was precluded. A fully automated solid-phase extraction (SPE) protocol, based on 96-well format plates, was used to isolate these compounds and furosemide (internal standard, IS) from plasma. Washing and elution steps were amended accordingly in order to minimize any matrix effect from components of the plasma without reducing the elution of the molecules of interest. The compounds were eluted from a C18 column and detected with an API 3000 triple-quadrupole mass spectrometer using negative electrospray ionization and multiple reaction monitoring (MRM). The assay was linear over the range 1.00-400 ng/mL for LOS and EXP-3174 and 0.500-200 ng/mL for HCTZ, respectively, when 200 mu l of plasma was used in the extraction. The overall intra- and interassay variations were within acceptance limits. The analysis time for each sample was 4 min, and more than 300 samples could be analyzed in one day by running the system overnight. The assay was simple, highly sensitive, selective, precise, fast, and it enables the reliable determination of LOS, EXP-3174 and HCTZ in pharmacokinetic or bioequivalence studies after per os administration of a single tablet containing both drugs.
引用
收藏
页码:593 / 601
页数:9
相关论文
共 34 条
[1]
SENSITIVE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF HYDROCHLOROTHIAZIDE IN HUMAN PLASMA [J].
AZUMAYA, CT .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1990, 532 (01) :168-174
[2]
Centre for Drug Evaluation and Research (CDER), 2001, GUID IND BIOAN METH
[3]
Determination of ketotifen and its conjugated metabolite in human plasma by liquid chromatography/tandem mass spectrometry: application to a pharmacokinetic study [J].
Chen, XY ;
Zhong, DF ;
Liu, D ;
Wang, YW ;
Han, Y ;
Gu, JK .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2003, 17 (22) :2459-2463
[4]
Efficacy and safety of angiotensin receptor blockers: A review of losartan in essential hypertension [J].
Conlin, PR .
CURRENT THERAPEUTIC RESEARCH-CLINICAL AND EXPERIMENTAL, 2001, 62 (02) :79-91
[5]
SIMPLE DETERMINATION OF HYDROCHLOROTHIAZIDE IN HUMAN PLASMA AND URINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
DEVRIES, JX ;
VOSS, A .
BIOMEDICAL CHROMATOGRAPHY, 1993, 7 (01) :12-14
[6]
Development of a rapid method for the determination of glimepiride in human plasma using liquid-liquid extraction based on 96-well format micro-tubes and liquid chromatography/tandem mass spectrometry [J].
Dotsikas, Y ;
Kousoulos, C ;
Tsatsou, G ;
Loukas, YL .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2005, 19 (14) :2055-2061
[7]
Development and validation of a rapid 96-well format based liquid-liquid extraction and liquid chromatography-tandem mass spectrometry analysis method for ondansetron in human plasma [J].
Dotsikas, Yannis ;
Kousoulos, Constantinos ;
Tsatsou, Georgia ;
Loukas, Yannis L. .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2006, 836 (1-2) :79-82
[8]
Simple high-performance liquid chromatographic method for determination of losartan and E-3174 metabolite in human plasma, urine and dialysate [J].
Farthing, D ;
Sica, D ;
Fakhry, I ;
Pedro, A ;
Gehr, TWB .
JOURNAL OF CHROMATOGRAPHY B, 1997, 704 (1-2) :374-378
[9]
Isolation of a 2:1 hydrochlorothiazide-formaldehyde adduct impurity in hydrochlorothiazide drug substance by preparative chromatography and characterization by electrospray ionization LC-MS [J].
Franolic, JD ;
Lehr, GJ ;
Barry, TL ;
Petzinger, G .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2001, 26 (04) :651-663
[10]
SIMULTANEOUS DETERMINATION OF A NOVEL ANGIOTENSIN-II RECEPTOR BLOCKING-AGENT, LOSARTAN, AND ITS METABOLITE IN HUMAN PLASMA AND URINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
FURTEK, CI ;
LO, MW .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1992, 573 (02) :295-301