Population patch clamp electrophysiology: a breakthrough technology for ion channel screening

被引:43
作者
Dale, Tim J. [1 ]
Townsend, Claire [1 ]
Hollands, Emma C. [1 ]
Trezise, Derek J. [1 ]
机构
[1] GlaxoSmithKline Inc, Res & Dev, Biol Reagents & Assay Dev, Harlow, Essex, England
关键词
D O I
10.1039/b706152h
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Population patch clamp (PPC) is a novel high throughput planar array electrophysiology technique that allows ionic currents to be recorded from populations of cells under voltage clamp. For the drug discovery pharmacologist, PPC promises greater speed and precision than existing methods for screening compounds at voltage-gated ion channel targets. Moreover, certain constitutively active or slow-ligand gated channels that have hitherto proved challenging to screen with planar array electrophysiology ( e. g. SK/IK channels) are now more accessible. In this article we review early findings using PPC and provide a perspective on its likely impact on ion channel drug discovery. To support this, we include some new data on ion channel assay duplexing and on modulator assays, approaches that have thus far not been described.
引用
收藏
页码:714 / 722
页数:9
相关论文
共 29 条
[1]  
Bridgland-Taylor M. H., 2006, Journal of Pharmacological and Toxicological Methods, V54, P189, DOI 10.1016/j.vascn.2006.02.003
[2]   Cloning, distribution and functional analysis of the type III sodium channel from human brain [J].
Chen, YH ;
Dale, TJ ;
Romanos, MA ;
Whitaker, WRJ ;
Xie, XM ;
Clare, JJ .
EUROPEAN JOURNAL OF NEUROSCIENCE, 2000, 12 (12) :4281-4289
[3]  
CLARE JJ, 2006, EXPRESSION ANAL RECO, pCH4
[4]   Identifying modulators of hERG channel activity using the PatchXpress® planar patch clamp [J].
Dubin, AE ;
Nasser, N ;
Rohrbacher, J ;
Hermans, AN ;
Marrannes, R ;
Grantham, C ;
Van Rossem, K ;
Cik, M ;
Chaplan, SR ;
Gallacher, D ;
Xu, J ;
Guia, A ;
Byrne, NG ;
Mathes, C .
JOURNAL OF BIOMOLECULAR SCREENING, 2005, 10 (02) :168-181
[5]   Automated ion channel screening:: patch clamping made easy [J].
Farre, Cecilia ;
Stoelzle, Sonja ;
Haarmann, Claudia ;
George, Michael ;
Brueggemann, Andrea ;
Fertig, Niels .
EXPERT OPINION ON THERAPEUTIC TARGETS, 2007, 11 (04) :557-565
[6]   Population patch clamp improves data consistency and success rates in the measurement of ionic currents [J].
Finkel, Alan ;
Wittel, Andrew ;
Yang, Naibo ;
Handran, Shawn ;
Hughes, Jan ;
Costantin, James .
JOURNAL OF BIOMOLECULAR SCREENING, 2006, 11 (05) :488-496
[7]   SMALL LINEAR CHLORIDE CHANNELS ARE ENDOGENOUS TO NONEPITHELIAL CELLS [J].
GABRIEL, SE ;
PRICE, EM ;
BOUCHER, RC ;
STUTTS, MJ .
AMERICAN JOURNAL OF PHYSIOLOGY, 1992, 263 (03) :C708-C713
[8]  
GONZALEZ JE, 2006, EXPRESSION ANAL RECO, pCH4
[9]  
GUO L, 2005, PHARM TOXICOL METHOD, V52, P123
[10]   A place for high-throughput electrophysiology in cardiac safety:: Screening hERG cell lines and novel compounds with the IonWorks HT™ system [J].
Guthrie, H ;
Livington, FS ;
Gubler, U ;
Garippa, R .
JOURNAL OF BIOMOLECULAR SCREENING, 2005, 10 (08) :832-840