Ion chromatographic analysis of tetracyclines using polymeric column and acidic eluent

被引:55
作者
Ding, XJ [1 ]
Mou, SF [1 ]
机构
[1] Chinese Acad Sci, Ecoenvironm Sci Res Ctr, Beijing 100085, Peoples R China
关键词
antibiotics; tetracyclines;
D O I
10.1016/S0021-9673(00)00779-2
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
High-performance ion chromatography (HPIC) is first successfully used to analyze tetracycline antibiotics (TCs) in this work. The TCs are well separated on a solvent compatible polymeric cation-exchange column within 12 min. Isocratic elution with acetonitrile-hydrochloride is very advantageous for routine analysis. HPIC may be seen as a specific variant of the more common high-performance liquid chromatography (HPLC) for water-soluble and polar pharmaceuticals with low hydrophobicity. The detection limits (signal-to-noise ratio similar or equal to3:1) of oxytetracycline (OTC), tetracycline (TC), chlortetracycline (CTC), doxycycline (DC) are 10, 10, 20 and 20 mug l(-1), respectively. Samples are prepared by vortex mixing with an ethylenediaminetetraacetic acid disodium salt (Na(2)EDTA)-McIlvaine buffer (pH 4.0) solution and the mixture filtrates through a molecular weight cut-off filter. The method has been successfully applied to monitor the OTC removal rate through every reactor in the process of OTC manufacturing wastewater treatment by bio-chemical technology. It is also applicable to determine the TCs residues in milk and milk powder with satisfying results. (C) 2000 Elsevier Science B.V. All rights reserved.