Accuracy of devices used for self-monitoring of blood glucose

被引:39
作者
Johnson, RN
Baker, JR
机构
[1] Green Lane Hosp, Dept Clin Biochem, Auckland 3, New Zealand
[2] Natl Womens Hosp, Auckland 3, New Zealand
关键词
near-patient testing; diabetes mellitus;
D O I
10.1177/000456329803500108
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
We measured the inaccuracy of 17 home blood glucose monitors (two visually read, eight colorimetric and seven amperometric). Using strips from single batch, blood glucose measurements were performed by three medical laboratory technologists on at least 50 capillary blood specimens from patients attending two. diabetes clinics. Additional capillary blood was deproteinized and assayed with hexokinase to give a whole blood glucose result. A dedicated glucose analyser was also studied to cross-validate the methodology. At a mean glucose concentration of about 9 mmol/L, monitor readings differed from the reference results by -5.1 to +19.5% with three systems failing to meet the American Diabetes Association guideline for total error of less than 15%. This problem would be alleviated by manufacturers adopting a common policy on calibration and on reporting as a plasma or whole blood value.
引用
收藏
页码:68 / 74
页数:7
相关论文
共 11 条
[1]  
[Anonymous], 1987, DIABETES CARE, V10, P95
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   WHAT IS BLOOD-GLUCOSE - CAN IT BE MEASURED [J].
BURRIN, JM ;
ALBERTI, KGMM .
DIABETIC MEDICINE, 1990, 7 (03) :199-206
[4]  
MOLITCH ME, 1994, DIABETES CARE, V17, P81
[5]  
*NAT COMM CLIN LAB, 1993, EP9T NCCLS
[6]  
NEESE JW, 1996, DHEW PUBL
[7]   EVALUATION OF A NEW BLOOD-GLUCOSE MONITORING-SYSTEM - IMPORTANCE OF THE APPROPRIATE REFERENCE METHOD [J].
PETERSON, CM ;
JOVANOVIC, L .
DIABETES CARE, 1984, 7 (06) :610-611
[8]  
RAINEY PM, 1995, AM J CLIN PATHOL, V103, P125
[9]   THE EFFECT OF INTENSIVE TREATMENT OF DIABETES ON THE DEVELOPMENT AND PROGRESSION OF LONG-TERM COMPLICATIONS IN INSULIN-DEPENDENT DIABETES-MELLITUS [J].
SHAMOON, H ;
DUFFY, H ;
FLEISCHER, N ;
ENGEL, S ;
SAENGER, P ;
STRELZYN, M ;
LITWAK, M ;
WYLIEROSETT, J ;
FARKASH, A ;
GEIGER, D ;
ENGEL, H ;
FLEISCHMAN, J ;
POMPI, D ;
GINSBERG, N ;
GLOVER, M ;
BRISMAN, M ;
WALKER, E ;
THOMASHUNIS, A ;
GONZALEZ, J ;
GENUTH, S ;
BROWN, E ;
DAHMS, W ;
PUGSLEY, P ;
MAYER, L ;
KERR, D ;
LANDAU, B ;
SINGERMAN, L ;
RICE, T ;
NOVAK, M ;
SMITHBREWER, S ;
MCCONNELL, J ;
DROTAR, D ;
WOODS, D ;
KATIRGI, B ;
LITVENE, M ;
BROWN, C ;
LUSK, M ;
CAMPBELL, R ;
LACKAYE, M ;
RICHARDSON, M ;
LEVY, B ;
CHANG, S ;
HEINHEINEMANN, M ;
BARRON, S ;
ASTOR, L ;
LEBECK, D ;
BRILLON, D ;
DIAMOND, B ;
VASILASDWOSKIN, A ;
LAURENZI, B .
NEW ENGLAND JOURNAL OF MEDICINE, 1993, 329 (14) :977-986
[10]   THE YELLOW-SPRINGS ANALYZER GIVES LOWER RESULTS THAN OTHER GLUCOSE-OXIDASE METHODS OF GLUCOSE MEASUREMENT [J].
UNWIN, N ;
SKILLEN, A ;
ALBERTI, KGMM .
ANNALS OF CLINICAL BIOCHEMISTRY, 1995, 32 :329-331