Development of response models for optimising HPLC methods

被引:34
作者
Dewé, W
Marini, RD
Chiap, P
Hubert, P
Crommen, J
Boulanger, B
机构
[1] Eli Lilly & Co, Lilly Dev Ctr, B-1348 Louvain, Belgium
[2] Univ Liege, CHU, Dept Pharm, B-4000 Cointe Ougree, Belgium
关键词
liquid chromatography; retention time; optimization; modeling; multi-criteria decision;
D O I
10.1016/j.chemolab.2004.04.016
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
For optimizing a liquid chromatographic method, direct modeling some responses like minimum resolution is sometimes difficult to perform. By modeling the retention time at apex and the times associated to 13.4% of the peak height, it is possible to obtain estimations of these responses using an adequate cascade of equations, these equations being the definitions of these responses. As a multi-criteria decision has to be taken in most of these optimizations, a Derringer's function can be defined for each response of interest and a overall desirability, on which the final is taken, can be derived. In addition, the chromatogram in any experimental point can be predicted. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:263 / 268
页数:6
相关论文
共 8 条
  • [1] Box G.E.P., 1960, Technometrics, V2, P455, DOI [10.1080/00401706.1960.10489912, DOI 10.1080/00401706.1960.10489912]
  • [2] *COUNC EUR, 2002, EUR PHARM ADD
  • [3] DERRINGER G, 1980, J QUAL TECHNOL, V12, P214, DOI 10.1080/00224065.1980.11980968
  • [4] LC method for the simultaneous determination of R-timolol and other closely related impurities in S-timolol maleate:: Optimization by use of an experimental design
    Marini, RD
    Chiap, P
    Boulanger, B
    Dewe, W
    Hubert, P
    Crommen, J
    [J]. JOURNAL OF SEPARATION SCIENCE, 2003, 26 (9-10) : 809 - 817
  • [5] Miller J. M., 1988, CHROMATOGRAPHY CONCE
  • [6] Neter J., 1990, APPL STAT MODELS
  • [7] SCHOENMAKERS P, 1986, OPTIMIZATON CHROMATO
  • [8] *US PHARM CONV, 2000, US PHARM NAT FORM, V19