Wear analysis of a retrieved hip implant with titanium nitride coating

被引:93
作者
Harman, MK [1 ]
Banks, SA [1 ]
Hodge, WA [1 ]
机构
[1] Good Samaritan Hosp, Orthopaed Res Lab, W Palm Beach, FL 33402 USA
关键词
titanium; titanium nitride; total hip arthroplasty; implant retrieval; wear; surface modification;
D O I
10.1016/S0883-5403(97)90164-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
There is increasing interest in using surface modification technology to improve the wear properties of titanium alloy and limit articular surface wear of metal and polyethylene components. This report details the in vivo wear performance of titanium nitride coating on a retrieved hip implant obtained postmortem from a low demand patient 1 year after total hip arthroplasty. Analysis of the well-functioning implant revealed that wear debris can originate from a titanium nitride coated femoral head, as delaminated surface asperities, and manifest as adhesive wear on the articular surface. The wear observed on this implant indicates that rigorous testing and evaluation of titanium nitride coating technology should be conducted prior to widespread use on total joint implants.
引用
收藏
页码:938 / 945
页数:8
相关论文
共 49 条
[1]   METALLIC WEAR IN FAILED TITANIUM-ALLOY TOTAL HIP REPLACEMENTS - A HISTOLOGICAL AND QUANTITATIVE-ANALYSIS [J].
AGINS, HJ ;
ALCOCK, NW ;
BANSAL, M ;
SALVATI, EA ;
WILSON, PD ;
PELLICCI, PM ;
BULLOUGH, PG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) :347-356
[2]   PRECISION-FIT SURFACE HEMIARTHROPLASTY FOR FEMORAL-HEAD OSTEONECROSIS - LONG-TERM RESULTS [J].
AMSTUTZ, HC ;
GRIGORIS, P ;
SAFRAN, MR ;
GRECULA, MJ ;
CAMPBELL, PA ;
SCHMALZRIED, TP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (03) :423-427
[3]  
[Anonymous], 1992, SOCIOLOGICA, V7
[4]  
Bankston A B, 1993, J Arthroplasty, V8, P315
[5]   THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[6]   METALLOSIS ASSOCIATED WITH A STABLE TITANIUM-ALLOY FEMORAL COMPONENT IN TOTAL HIP-REPLACEMENT - A CASE-REPORT [J].
BLACK, J ;
SHERK, H ;
BONINI, J ;
ROSTOKER, WR ;
SCHAJOWICZ, F ;
GALANTE, JO .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (01) :126-130
[7]  
BLOEBAUM RD, 1993, J ARTHROPLASTY, V8, P195, DOI 10.1016/S0883-5403(09)80013-2
[8]  
BUECHEL FF, 1993, 8 ANN JOINT REPL S P
[9]  
BUECHEL FF, 1994, 9 ANN JOINT REPL S P
[10]  
BUECHEL FF, 1993, BUECHELPAPPAS INTEGR