MR-guided interstitial cryotherapy of the liver with a novel, nitrogen-cooled cryoprobe

被引:39
作者
Tacke, J
Adam, G
Speetzen, R
Brucksch, K
Bücker, A
Heschel, I
Prescher, A
van Vaals, JJ
Hunter, DW
Rau, G
Günther, RW
机构
[1] Aachen Tech Univ, Dept Diagnost Radiol, D-52074 Aachen, Germany
[2] Aachen Tech Univ, Helmholtz Inst Biomed Engn, D-52074 Aachen, Germany
[3] Aachen Tech Univ, Dept Anat, D-52074 Aachen, Germany
[4] Philips, Best, Netherlands
[5] Fairview Univ, Med Ctr, Dept Radiol, Minneapolis, MN USA
关键词
cryotherapy; liquid nitrogen; interventional MR; interstitial therapy; liver;
D O I
10.1002/mrm.1910390304
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of the study was to test a newly developed, MR-compatible, liquid nitrogen-cooled cryoprobe. The probe has an outer diameter of 3.5 mm and was specifically developed for percutaneous, MR-guided, interstitial cryotherapy. The probe was inserted percutaneously into the livers of 10 rabbits. The cryotherapy procedure was monitored with a surface coil in a 1.5 Tesla magnet using a gradient echo sequence. Follow-up examinations were performed 3 and 7 days after the freezing procedure using T-1- and T-2-weighted spin echo sequences. At 7 days the animals were sacrificed and the cryolesions were examined histologically. The cryoprobe enabled artifact-free MR imaging of the "ice-ball" formation during freezing of the rabbit liver. After 1 min of freezing, the iceball at the tip of the probe showed an average maximum diameter of 10.8 mm. No bleeding complications were observed during or after the freezing procedure. Histologic examination 7 days after cryotherapy confirmed that the Liver lesions were the same size as had been predicted by the images of the acute iceball. This new, percutaneously inserted, MR-compatible, liquid-nitrogen cooled cryoprobe allows accurate, artifact-free MR imaging of interstitial cryotherapy.
引用
收藏
页码:354 / 360
页数:7
相关论文
共 29 条
[1]  
Bird HM, 1949, ANAESTHESIA, V4, P10
[2]   CRYOTHERAPY FOR LIVER METASTASES - A NEW APPROACH [J].
CHARNLEY, RM ;
DORAN, J ;
MORRIS, DL .
BRITISH JOURNAL OF SURGERY, 1989, 76 (10) :1040-1041
[4]   RESULTS OF CRYOTHERAPY FOR EYELID MALIGNANCIES [J].
FRAUNFELDER, FT ;
ZACARIAN, SA ;
WINGFIELD, DL ;
LIMMER, BL .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1984, 97 (02) :184-188
[5]   REAL-TIME ULTRASONIC MONITORING OF HEPATIC CRYOSURGERY [J].
GILBERT, JC ;
ONIK, GM ;
HODDICK, WK ;
RUBINSKY, B .
CRYOBIOLOGY, 1985, 22 (04) :319-330
[6]   MRI-MONITORED CRYOSURGERY IN THE RABBIT BRAIN [J].
GILBERT, JC ;
RUBINSKY, B ;
ROOS, MS ;
WONG, STS ;
BRENNAN, KM .
MAGNETIC RESONANCE IMAGING, 1993, 11 (08) :1155-1164
[7]   THE OPERATION AND EFFICACY OF CRYOSURGICAL, NITROUS OXIDE-DRIVEN CRYOPROBE .1. CRYOPROBE PHYSICAL CHARACTERISTICS - THEIR EFFECTS ON CELL CRYODESTRUCTION [J].
HOMASSON, JP ;
THIERY, JP ;
ANGEBAULT, M ;
OVTRACHT, L ;
MAIWAND, O .
CRYOBIOLOGY, 1994, 31 (03) :290-304
[8]  
ISHIHARA Y, 1997, P ISMRM 5 ANN M VANC, P1953
[9]  
KAHN T, 1997, P ISMRM 5 ANN M VANC, P528
[10]  
KUFLIK EG, 1990, CRYOSURGICAL TREATME, P243