A modified semi-automated technique for in vitro assessment of encrustation of materials

被引:5
作者
Gepi-Attee, S [1 ]
Feneley, RCL [1 ]
机构
[1] Southmead Hosp, Bristol Urol Inst, Dept Urol, Bristol, Avon, England
关键词
encrustation; alloplast; artificial urine; struvite; hydroxyapatite;
D O I
10.1243/0954411981534592
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In vitro study of encrustation is an important part of assessment of materials as potential alloplasts or devices in the urinary tract. This modified semi-automated technique comprises a circular reaction chamber with an encrustation mixture, the level of which is controlled by a float switch which operates the exit peristaltic pump. The composition of the reactants used simulates infected urine with alkaline pH. Results of a preliminary study of the deposits by scanning electron micrography (SEM) and energy dispersive X-ray (EDX) microanalysis are consistent with struvite and hydroxyapatite, similar to the main minerals deposited on urinary catheters. It is a relatively simple, effective and inexpensive set-up for study of encrustation on materials.
引用
收藏
页码:475 / 478
页数:4
相关论文
共 8 条
[1]  
BURNS JR, 1980, INVEST UROL, V18, P167
[2]  
Cox A J, 1987, Eng Med, V16, P37
[3]   MINERALOGICAL STUDIES OF URINE - THE RELATIONSHIP OF APATITE, BRUSHITE AND STRUVITE TO URINARY PH [J].
ELLIOT, JS ;
QUAIDE, WL ;
SHARP, RF ;
LEWIS, L .
JOURNAL OF UROLOGY, 1958, 80 (04) :269-271
[4]  
FINLAYSON B, 1973, INVEST UROL, V10, P429
[5]  
GLEESON MJ, 1989, T AM SOC ART INT ORG, V35, P495
[6]  
MILLER JM, 1974, INVEST UROL, V11, P402
[7]  
ROBRIQUES G, 1979, ARCH PHYS MED REHAB, V60, P266
[8]   ENCRUSTATION OF CATHETER MATERIALS IN-VITRO [J].
SRINIVASAN, V ;
CLARK, SS .
JOURNAL OF UROLOGY, 1972, 108 (03) :473-+