Constructing a new optical sensor for monitoring ammonia in water samples using bis(acetylacetoneethylendiamine)-tributylphosphin cobalt(III) tetraphenylborate complex-coated triacetylcellulose

被引:21
作者
Absalan, G [1 ]
Soleimani, M [1 ]
Asadi, M [1 ]
Ahmadi, MB [1 ]
机构
[1] Shiraz Univ, Dept Chem, Shiraz 71454, Iran
关键词
D O I
10.2116/analsci.20.1433
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new ammonia optical sensor was designed using bis(acetylacetoneethylendiamine)tributylphosphin cobalt(III) tetraphenylborate complex, coated on transparent triacety1cellulose film as membrane. The change in the absorbance of the optode at the maximum wavelength of 408 nm was related to ammonia concentration in aqueous samples. A buffer solution with a pH of 9 (sodium borate-HCl) was used. The optode was fully regenerated in pH 2. The linear dynamic range for determination of ammonia was 3.3 x 10(-4) to 6.9 x 10(-1) mol l(-1) with a detection limit of 5.0 x 10(-5) Mol l(-1) and a response time range of 4 - 6 min. This membrane was successfully applied for determination of ammonia in drinking water.
引用
收藏
页码:1433 / 1436
页数:4
相关论文
共 41 条
[1]   AN OPTICAL POTASSIUM-ION SENSOR [J].
ALDER, JF ;
ASHWORTH, DC ;
NARAYANASWAMY, R ;
MOSS, RE ;
SUTHERLAND, IO .
ANALYST, 1987, 112 (08) :1191-1192
[2]   Real-time monitoring of ammonia-oxidizing activity in a nitrifying biofilm by amoA mRNA analysis [J].
Aoi, Y ;
Shiramasa, Y ;
Tsuneda, S ;
Hirata, A ;
Kitayama, A ;
Nagamune, T .
WATER SCIENCE AND TECHNOLOGY, 2002, 46 (1-2) :439-442
[3]   AN OPTICAL CALCIUM-ION SENSOR [J].
ASHWORTH, DC ;
HUANG, HP ;
NARAYANASWAMY, R .
ANALYTICA CHIMICA ACTA, 1988, 213 (1-2) :251-257
[4]  
BARNES AR, 1990, PHARM ACTA HELV, V65, P258
[5]   A NEW ION SENSOR BASED ON FIBER OPTICS [J].
BRIGHT, FV ;
POIRIER, GE ;
HIEFTJE, GM .
TALANTA, 1988, 35 (02) :113-118
[6]  
CAREY WP, 1989, ANAL CHEM, V61, P1647
[7]   FIBER-OPTIC TIME-RESOLVED FLUORESCENCE SENSOR FOR THE SIMULTANEOUS DETERMINATION OF AL-3+ AND GA-3+ OR IN-3+ [J].
CARROLL, MK ;
BRIGHT, FV ;
HIEFTJE, GM .
ANALYTICAL CHEMISTRY, 1989, 61 (15) :1768-1772
[8]   REVERSIBLE DETECTION OF SULFUR-DIOXIDE BY USING A MULTIPLE REFLECTING OPTICAL WAVE-GUIDE SENSOR [J].
COOK, RL ;
MACDUFF, RC ;
SAMMELLS, AF .
ANALYTICA CHIMICA ACTA, 1989, 226 (01) :153-158
[9]   DIRECT MEASUREMENT OF AMMONIA IN AMBIENT AIR [J].
DAVID, DJ ;
WILLSON, MC ;
RUFFIN, DS .
ANALYTICAL LETTERS, 1976, 9 (04) :389-404
[10]  
DIKERT FL, 1989, ANAL CHEM, V61, P2306