Whole blood international normalization ratio measurements in children using near-patient monitors

被引:27
作者
Nowatzke, WL
Landt, M
Smith, C
Wilhite, T
Canter, C
Luchtman-Jones, L
机构
[1] Washington Univ, Sch Med, Core Lab Clin Studies, Dept Pediat, St Louis, MO 63110 USA
[2] Washington Univ, Sch Med, Dept Pathol, St Louis, MO 63110 USA
[3] St Louis Childrens Hosp, St Louis, MO 63178 USA
关键词
international normalized ratio; pediatrics; point-of-care;
D O I
10.1097/00043426-200301000-00008
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To report a comparison of international normalization ratio (INR) measurements on four near-patient (point-of-care or bedside) whole blood INR monitors in children. Patients and Methods: The INR results from 19 ambulatory pediatric subjects (30 hospital visits) receiving warfarin sodium were analyzed on four near-patient monitors and compared with plasma INR measurements on the laboratory CA-1000 Analyze. The instruments evaluated were CoaguChek, Hemochron Jr. Signature, ProTime Microcoagulation System, and RapidpointCoag. Results: The INR measurements ranged from 1.05 to 5.25. Over the entire INR range, the near-patient instrument with the least bias relative to the CA-1000 was the RapidpointCoag (r(2) = 0.923). The correlations (r(2)) of the CoaguCheck, Hemochron Jr., and ProTime were 0.877, 0.834, and 0.885, respectively. Precision studies involved repeated analysis of one nonmedicated adult (mean CA-1000 INR = 0.908) and one adult receiving oral anticoagulation therapy (mean CA-1000 INR = 2.42). The coefficient of variation on the near-patient monitors for both adult volunteers ranged from 4.9% to 22.3%. Bilirubin levels up to 20 mg/dL did not interfere in any of the methods. Conclusions: Near-patient testing whole blood INR monitors offer acceptably accurate and precise measurements. Values obtained on near-patient monitors may vary considerably from the reference method, and data obtained should serve as a supplement to, but not a replacement for, routine clinical laboratory measurements.
引用
收藏
页码:33 / 37
页数:5
相关论文
共 13 条
[1]  
ANDREW M, 1994, THROMB HAEMOSTASIS, V71, P265
[2]   The prevention and treatment of thromboembolic disease in children: A need for thrombophilia programs [J].
Andrew, M ;
Michelson, AD ;
Bovill, T ;
Leaker, M ;
Massicotte, P ;
Marzinotto, V ;
Brooker, LA .
JOURNAL OF PEDIATRIC HEMATOLOGY ONCOLOGY, 1997, 19 (01) :7-22
[3]  
Andrew M., 1996, Blood, V88, p179A
[4]  
ANDREW M, 1995, BLOOD S, V86, pA864
[5]   PATIENT SELF-MANAGEMENT OF ORAL ANTICOAGULATION GUIDED BY CAPILLARY (FINGERSTICK) WHOLE-BLOOD PROTHROMBIN TIMES [J].
ANSELL, J ;
HOLDEN, A ;
KNAPIC, N .
ARCHIVES OF INTERNAL MEDICINE, 1989, 149 (11) :2509-2511
[6]   A comparison of point-of-gate instruments designed for monitoring oral anticoagulation with standard laboratory methods [J].
Gosselin, R ;
Owings, JT ;
White, RH ;
Hutchinson, R ;
Branch, J ;
Mahackian, K ;
Johnston, M ;
Larkin, EC .
THROMBOSIS AND HAEMOSTASIS, 2000, 83 (05) :698-703
[7]   MECHANISM OF ACTION, CLINICAL EFFECTIVENESS, AND OPTIMAL THERAPEUTIC RANGE [J].
HIRSH, J ;
DALEN, JE ;
DEYKIN, D ;
POLLER, L ;
BUSSEY, H .
CHEST, 1995, 108 (04) :S231-S246
[8]  
Horstkotte D., 1996, European Heart Journal, V17, P230
[9]  
HORSTKOTTE D, 1998, J THROMB THROMBOLYS, V5, pS19
[10]   CURRENT CONCEPTS IN ANTICOAGULANT-THERAPY [J].
LITIN, SC ;
GASTINEAU, DA .
MAYO CLINIC PROCEEDINGS, 1995, 70 (03) :266-272