Effect of sensory substitution on suture-manipulation forces for robotic surgical systems

被引:209
作者
Kitagawa, M
Dokko, D
Okamura, AM
Yuh, DD
机构
[1] Johns Hopkins Univ, Dept Mech Engn, Baltimore, MD 21218 USA
[2] Johns Hopkins Med Inst, Div Cardiac Surg, Baltimore, MD 21205 USA
关键词
D O I
10.1016/j.jtcvs.2004.05.029
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: Direct haptic (force or tactile) feedback is not yet available in commercial robotic surgical systems. Previous work by our group and others suggests that haptic feedback might significantly enhance the execution of surgical tasks requiring fine suture manipulation, specifically those encountered in cardiothoracic surgery. We studied the effects of substituting direct haptic feedback with visual and auditory cues to provide the operating surgeon with a representation of the forces he or she is applying with robotic telemanipulators. Methods: Using the robotic da Vinci Surgical system (Intuitive Surgical, Inc, Sunnyvale, Calif), we compared applied forces during a standardized surgical knot-tying task under 4 different sensory-substitution scenarios: no feedback, auditory feedback, visual feedback, and combined auditory-visual feedback. Results: The forces applied with these sensory-substitution modes more closely approximate suture tensions achieved under ideal haptic conditions (ie, hand ties) than forces applied without such sensory feedback. The consistency of applied forces during robot-assisted suture tying aided by visual feedback or combined auditory-visual feedback sensory substitution is superior to that achieved with hand ties. Robot-assisted ties aided with auditory feedback revealed levels of consistency that were generally equivalent or superior to those attained with hand ties. Visual feedback and auditory feedback improve the consistency of robotically applied forces. Conclusions: Sensory substitution, in the form of visual feedback, auditory feedback, or both, confers quantifiable advantages in applied force accuracy and consistency during the performance of a simple surgical task.
引用
收藏
页码:151 / 158
页数:8
相关论文
共 23 条
[1]   The Leipzig experience with robotic valve surgery [J].
Autschbach, R ;
Onnasch, JF ;
Falk, V ;
Walther, T ;
Krüger, M ;
Schilling, LO ;
Mohr, FW .
JOURNAL OF CARDIAC SURGERY, 2000, 15 (01) :82-87
[2]   Closed-chest coronary artery bypass grafting on the beating heart with the use of a computer-enhanced surgical robotic system [J].
Boyd, WD ;
Rayman, R ;
Desai, ND ;
Menkis, AH ;
Dobkowski, W ;
Ganapathy, S ;
Kiaii, B ;
Jablonsky, G ;
McKenzie, FN ;
Novick, RJ .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2000, 120 (04) :807-809
[3]   Minimally invasive videoscopic mitral valve surgery: The current role of surgical robotics [J].
Chitwood, WR ;
Nifong, LW .
JOURNAL OF CARDIAC SURGERY, 2000, 15 (01) :61-75
[4]  
CHITWOOD WR, 2003, HEART SURG FORUM, V6, P106
[5]   Robotic cardiac surgery: Present and future applications [J].
Czibik, G ;
D'Ancona, G ;
Donias, HW ;
Karamanoukian, HL .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2002, 16 (04) :495-501
[6]   Endoscopic coronary artery bypass grafting - The first steps on a long journey [J].
Damiano, RJ .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2000, 120 (04) :806-807
[7]  
DEBUS T, 2001, P SPIE INT SOC OPT E, V4570
[8]   Haptic force-feedback devices for the office computer: Performance and musculoskeletal loading issues [J].
Dennerlein, JT ;
Yang, MC .
HUMAN FACTORS, 2001, 43 (02) :278-286
[9]  
DENNERLEIN JT, 2000, P C HUM FACT COMP SY
[10]   Robotically assisted left ventricular epicardial lead implantation for biventricular pacing [J].
DeRose, JJ ;
Ashton, RC ;
Belsley, S ;
Swistel, DG ;
Vloka, M ;
Ehlert, F ;
Shaw, R ;
Sackner-Bernstein, J ;
Hillel, Z ;
Steinberg, JS .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 41 (08) :1414-1419