A novel RF coil configuration for in-vivo and ex-vivo imaging of arthritic rabbit knee joints

被引:7
作者
Anastasiou, A [1 ]
Hall, LD [1 ]
机构
[1] Univ Cambridge, Sch Clin Med, Herchel Smith Lab Med Chem, Cambridge CB2 2PZ, England
关键词
MRI; rabbit; articular cartilage; decoupling;
D O I
10.1016/S0730-725X(02)00636-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to design and build an optimized Radio Frequency (RF) coil configuration, that would facilitate the acquisition of high resolution 3-dimensional (3D) images of arthritic and normal rabbit knees. A surface coil transmit surface coil receive configuration was built, in order to ensure adequate B, homogeneity over the imaging volume and maximum filling factor, and hence to maximize the Signal to Noise ratio (SNR) and resolution of the 3-dimensional images. The two coils were passively decoupled using crossed diodes and lambda/4 lines, both during the transmit and receive phases of the imaging experiment. A specialized animal bed, to optimize the use of the coils and minimize the experiment setup time was designed and constructed. Three dimensional images of resolution 156 X 156 X 468 mum, were acquired in 20 min; the results, in terms both of the high resolution images and the ease with which the experimental setup could be reproduced, demonstrated that this configuration is ideal for imaging rabbit knee joints. (C) 2003 Elsevier Science Inc. All rights reserved.
引用
收藏
页码:61 / 68
页数:8
相关论文
共 26 条
[1]   MRI contrast enhanced study of cartilage proteoglycan degradation in the rabbit knee [J].
Bacic, G ;
Liu, KJ ;
Goda, F ;
Hoopes, PJ ;
Rosen, GM ;
Swartz, HM .
MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (05) :764-768
[2]   ELIMINATION OF COUPLING BETWEEN CYLINDRICAL TRANSMIT COILS AND SURFACE-RECEIVE COILS FOR INVIVO NMR [J].
BENDALL, MR ;
CONNELLY, A ;
MCKENDRY, JM .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (01) :157-163
[3]   DEPTH-RESOLVED SURFACE-COIL SPECTROSCOPY (DRESS) FOR INVIVO H-1, P-31, AND C-13 NMR [J].
BOTTOMLEY, PA ;
FOSTER, TB ;
DARROW, RD .
JOURNAL OF MAGNETIC RESONANCE, 1984, 59 (02) :338-342
[4]   High-resolution MRI detects cartilage swelling at the early stages of experimental osteoarthritis [J].
Calvo, E ;
Palacios, I ;
Delgado, E ;
Ruiz-Cabello, J ;
Hernández, P ;
Sánchez-Pernaute, O ;
Egido, J ;
Herrero-Beaumont, G .
OSTEOARTHRITIS AND CARTILAGE, 2001, 9 (05) :463-472
[5]   HIGH-RESOLUTION NMR IMAGING OF AN ANTIGEN-INDUCED ARTHRITIS IN THE RABBIT KNEE [J].
CHECKLEY, D ;
JOHNSTONE, D ;
TAYLOR, K ;
WATERTON, JC .
MAGNETIC RESONANCE IN MEDICINE, 1989, 11 (02) :221-235
[6]  
Dawson J, 1999, ARTHRITIS RHEUM, V42, P119, DOI 10.1002/1529-0131(199901)42:1<119::AID-ANR15>3.0.CO
[7]  
2-A
[8]  
Duffin W. J., 1990, ELECT MAGNETISM
[9]   ELECTRONIC DECOUPLING OF SURFACE-COIL RECEIVERS FOR NMR IMAGING AND SPECTROSCOPY [J].
EDELSTEIN, WA ;
HARDY, CJ ;
MUELLER, OM .
JOURNAL OF MAGNETIC RESONANCE, 1986, 67 (01) :156-161
[10]  
Fukushima E., 1981, EXPT PULSE NMR