Clinical evaluation of the Non-invasive Cardiac Output (NICO) monitor in the intensive care unit

被引:52
作者
van Heerden, PV
Baker, S
Lim, SI
Weidman, C
Bulsara, M
机构
[1] Sir Charles Gairdner Hosp, Dept Intens Care, Perth, WA, Australia
[2] Sir Charles Gairdner Hosp, Dept Pharmacol, Perth, WA, Australia
关键词
monitoring : cardiac output; non-invasive;
D O I
10.1177/0310057X0002800412
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The Non-invasive Cardiac Output (NICO) monitor (Novametrix Medical Systems Inc., Wallingford, CT, U.S.A.) utilizes a minimally-invasive partial rebreathing methods to determine cardiac output by means of a differential form of the Fick equation. We evaluated the NICO monitor by comparing its output to paired measurements obtained by the standard thermodilution (TD) technique in patients who had recently undergone cardiac surgery. Forty-two paired measurements were carried out in 12 patients. The correlation between the two methods was moderate with a correlation coefficient of 0.691. Repeated measures ANOVA showed that TD measures of cardiac output were significantly higher than those obtained by the NICO monitor (P=0.0003). Comparison of the two techniques using the methods described by Bland and Altman showed decreased correlation at higher values of cardiac output. We conclude that the NICO monitor may well have a place in intensive care monitoring, provided patients are not breathing spontaneously and are able to tolerate a 4 mmHg rise in PaCO2. It is less suitable for use in patients with a high cardiac output state.
引用
收藏
页码:427 / 430
页数:4
相关论文
共 7 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   NONINVASIVE MEASUREMENT OF CARDIAC-OUTPUT USING PARTIAL CO2 REBREATHING [J].
CAPEK, JM ;
ROY, RJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (09) :653-661
[3]   Partial carbon dioxide rebreathing: A reliable technique for noninvasive measurement of nonshunted pulmonary capillary blood flow [J].
deAbreu, MG ;
Quintel, M ;
Ragaller, M ;
Albrecht, DM .
CRITICAL CARE MEDICINE, 1997, 25 (04) :675-683
[4]   Clinical evaluation of a new noninvasive method of cardiac output measurement - Preliminary results in CABG patients [J].
Guzzi, L ;
Jaffe, MB ;
Orr, JA .
ANESTHESIOLOGY, 1998, 89 (3A) :U409-U409
[5]  
HARYADI DG, 1998, ANESTHESIOLOGY, V89, pA534
[6]  
Jopling MW, 1998, ANESTHESIOLOGY, V89, pU410
[7]   Evaluation of partial re-breathing cardiac output measurement during surgery [J].
Kuck, K ;
Haryadi, DG ;
Orr, JA ;
Bailey, PL .
ANESTHESIOLOGY, 1998, 89 (3A) :U409-U409