High-resolution diffusion imaging using phase-corrected segmented echo-planar imaging

被引:26
作者
Brockstedt, S [1 ]
Moore, JR
Thomsen, C
Holtås, S
Ståhlberg, F
机构
[1] Univ Lund Hosp, Dept Diagnost Radiol, S-22185 Lund, Sweden
[2] Siemens Med Syst Inc, Iselin, NJ 08830 USA
[3] Univ Copenhagen, Rigshosp, Dept Diagnost Radiol, DK-1168 Copenhagen, Denmark
[4] Univ Lund Hosp, Dept Radiat Phys, S-22185 Lund, Sweden
关键词
diffusion imaging; high resolution; phase corrections; segmented EPI; ischemic stroke;
D O I
10.1016/S0730-725X(00)00159-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Diffusion magnetic resonance imaging (MRI) was performed with a high-resolution segmented echo-planar imaging technique, which provided images with substantially less susceptibility artifacts than images obtained with single-shot echo-planar imaging (EPI). Diffusion imaging performed with any multishot pulse sequence is inherently sensitive to motion artifacts and in order to reduce motion artifacts, the presented method utilizes navigator echo phase corrections, performed after a one-dimensional Fourier transform along the frequency-encoding direction. Navigator echo phases were fitted to a straight line prior to phase correction to avoid errors from internal motion. In vivo imaging was performed using electro cardiographic (ECG) triggering. Apparent diffusion coefficient (ADC) maps were calculated on a pixel-by-pixel basis using up to seven diffusion sensitivities, ranging from b = 0 to 1129 x 10(6) s/m(2). (C) 2000 Elsevier Science Inc. All rights reserved.
引用
收藏
页码:649 / 657
页数:9
相关论文
共 22 条
[1]   ANALYSIS AND CORRECTION OF MOTION ARTIFACTS IN-DIFFUSION WEIGHTED IMAGING [J].
ANDERSON, AW ;
GORE, JC .
MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (03) :379-387
[2]   Quantitative diffusion coefficient maps using fast spin-echo MRI [J].
Brockstedt, S ;
Thomsen, C ;
Wirestam, R ;
Holtås, S ;
Ståhlberg, F .
MAGNETIC RESONANCE IMAGING, 1998, 16 (08) :877-886
[3]   IMAGE-RECONSTRUCTION FOR ECHO PLANAR IMAGING WITH NONEQUIDISTANT KAPPA-SPACE SAMPLING [J].
BRUDER, H ;
FISCHER, H ;
REINFELDER, HE ;
SCHMITT, F .
MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (02) :311-323
[4]   Diffusion-weighted interleaved echo-planar imaging with a pair of orthogonal navigator echoes [J].
Butts, K ;
deCrespigny, A ;
Pauly, JM ;
Moseley, M .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (05) :763-770
[5]   Isotropic diffusion-weighted and spiral-navigated interleaved EPI for routine imaging of acute stroke [J].
Butts, K ;
Pauly, J ;
deCrespigny, A ;
Moseley, M .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (05) :741-749
[6]   INTERLEAVED ECHO-PLANAR IMAGING ON A STANDARD MRI SYSTEM [J].
BUTTS, K ;
RIEDERER, SJ ;
EHMAN, RL ;
THOMPSON, RM ;
JACK, CR .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (01) :67-72
[7]   ECHO-PLANAR IMAGING OF THE LIVER WITH A STANDARD MR-IMAGING SYSTEM [J].
BUTTS, K ;
RIEDERER, SJ ;
EHMAN, RL ;
FELMLEE, JP ;
GRIMM, RC .
RADIOLOGY, 1993, 189 (01) :259-264
[8]  
BUTTS K, 1997, P 5 SCI M INT SOC MA, P222
[9]   ADAPTIVE TECHNIQUE FOR HIGH-DEFINITION MR IMAGING OF MOVING STRUCTURES [J].
EHMAN, RL ;
FELMLEE, JP .
RADIOLOGY, 1989, 173 (01) :255-263
[10]   Cerebrospinal fluid-suppressed high-resolution diffusion imaging of human brain [J].
Falconer, JC ;