An improved biomechanical testing protocol for evaluating multilevel cervical instrumentation in a human cadaveric corpectomy model

被引:2
作者
DiAngelo, DJ [1 ]
Foley, KT [1 ]
机构
[1] Univ Tennessee, Ctr Hlth Sci, Sch Biomed Engn, Memphis, TN 38163 USA
来源
SPINAL IMPLANTS: ARE WE EVALUATING THEM APPROPRIATELY | 2003年 / 1431卷
关键词
biomechanical testing; spinal instrumentation; corpectomy; cervical spine; biomechanics; anterior cervical plate; instrumented strut-graft mechanics;
D O I
10.1520/STP11142S
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Spinal instrumentation is commonly used to stabilize the multi-level strut-grafted cervical spine. There are no standard tissue-based testing protocols for evaluating spinal devices. An improved biomechanical testing protocol was developed to study the strut-graft mechanics of the instrumented cervical spine in flexion and extension. The biomechanical stability of two different anterior cervical plating systems were evaluated and compared to the harvested condition. A force sensing strut-graft (fssg) was used to measure the axial compressive load. Parameters of stiffness, segmental vertebral motion, and strut-graft loads were statistically compared with a one-way anova (p<0.05) to determine differences between the spine conditions. Applying a bending moment distribution across the cervical spine resulted in a motion response that closely matched the in vivo case. Fssg loads were affected by plate application and specific plate design features. The testing protocol has been used to study the biomechanical stability of various multi-level strut-graft cervical instrumentation. Information of strut-graft loads aided in better understanding graft-plate load-sharing mechanics and clinically-observed multi-level instrumentation failure mechanisms.
引用
收藏
页码:155 / 172
页数:18
相关论文
共 37 条
[1]
INSTANTANEOUS AXES OF ROTATION OF THE TYPICAL CERVICAL MOTION SEGMENTS .1. AN EMPIRICAL-STUDY OF TECHNICAL ERRORS [J].
AMEVO, B ;
MACINTOSH, JE ;
WORTH, D ;
BOGDUK, N .
CLINICAL BIOMECHANICS, 1991, 6 (01) :31-37
[2]
INSTANTANEOUS AXES OF ROTATION OF THE TYPICAL CERVICAL MOTION SEGMENTS - A STUDY IN NORMAL VOLUNTEERS [J].
AMEVO, B ;
WORTH, D ;
BOGDUK, N .
CLINICAL BIOMECHANICS, 1991, 6 (02) :111-117
[3]
CHEN J, 1996, THESIS UT MEMPHIS
[4]
Load-sharing characteristics of stabilized lumbar spine segments [J].
Cripton, PA ;
Jain, GM ;
Wittenberg, RH ;
Nolte, LP .
SPINE, 2000, 25 (02) :170-179
[5]
CRIPTON PA, P 2001 CERV SPIN RES, P214
[6]
Anterior cervical plating reverses load transfer through multilevel strut-grafts [J].
DiAngelo, DJ ;
Foley, KT ;
Vossel, KA ;
Rampersaud, YR ;
Jansen, TH .
SPINE, 2000, 25 (07) :783-795
[7]
DIANGELO DJ, 1998, 1998 ASME ADV BIOENG
[8]
DIANGELO DJ, P 2000 CERV SPIN RES, P203
[9]
DIANGELO DJ, 1997, 1997 ASME ADV BIOENG
[10]
CERVICAL-SPINE MOTION IN THE SAGITTAL PLANE - KINEMATIC AND GEOMETRIC PARAMETERS [J].
DIMNET, J ;
PASQUET, A ;
KRAG, MH ;
PANJABI, MM .
JOURNAL OF BIOMECHANICS, 1982, 15 (12) :959-969