Subnanometric measurements of evanescent wave penetration depth using total internal reflection microscopy combined with fluorescent correlation spectroscopy

被引:25
作者
Harlepp, S
Robert, J
Darnton, NC
Chatenay, D
机构
[1] ULP, CNRS, Lab Dynam Fluides Complexes, F-67000 Strasbourg, France
[2] Univ Chicago, Inst Biophys Dynam, Chicago, IL 60637 USA
[3] Univ Chicago, James Franck Inst, Chicago, IL 60637 USA
[4] Harvard Univ, Rowland Inst, Cambridge, MA 02142 USA
关键词
D O I
10.1063/1.1802374
中图分类号
O59 [应用物理学];
学科分类号
摘要
In total internal reflection microscopy (TIRM), quantitative interpretation of results often requires a precise knowledge of the penetration depth of the evanescent wave. Standard TIRM practice is to calculate this depth from the microscope's geometry, but this can introduce significant errors. We show how to calibrate the penetration depth of an evanescent wave in TIRM. An evanescent wave is obtained by illuminating a surface at an incident angle greater than the critical angle. Its penetration depth generally depends on the wavelength and the incident angle of the illumination, and on the indices of refraction on either side of the reflecting suface, but cannot be larger than the field of view. By introducing a fluorescent species (such as fluorescein) and measuring its diffusion time, it is possible to measure very precisely the penetration depth of the evanescent wave. (C) 2004 American Institute of Physics.
引用
收藏
页码:3917 / 3919
页数:3
相关论文
共 14 条
[1]  
HARLEPP S, UNPUB
[2]   Focal volume optics and experimental artifacts in confocal fluorescence correlation spectroscopy [J].
Hess, ST ;
Webb, WW .
BIOPHYSICAL JOURNAL, 2002, 83 (04) :2300-2317
[3]   Measurement of colloidal forces with TIRM. [J].
Prieve, DC .
ADVANCES IN COLLOID AND INTERFACE SCIENCE, 1999, 82 (1-3) :93-125
[4]  
RIGLER R, SPRINGER SERIES
[5]   Molecular dynamics in living cells observed by fluorescence correlation spectroscopy with one- and two-photon excitation [J].
Schwille, P ;
Haupts, U ;
Maiti, S ;
Webb, WW .
BIOPHYSICAL JOURNAL, 1999, 77 (04) :2251-2265
[6]  
Schwille P, 1999, CYTOMETRY, V36, P176, DOI 10.1002/(SICI)1097-0320(19990701)36:3<176::AID-CYTO5>3.0.CO
[7]  
2-F
[8]   Total internal reflection with fluorescence correlation spectroscopy: Combined surface reaction and solution diffusion [J].
Starr, TE ;
Thompson, NL .
BIOPHYSICAL JOURNAL, 2001, 80 (03) :1575-1584
[9]   Variable-angle total internal reflection fluorescence microscopy (VA-TIRFM): realization and application of a compact illumination device [J].
Stock, K ;
Sailer, R ;
Strauss, WSL ;
Lyttek, M ;
Steiner, R ;
Schneckenburger, H .
JOURNAL OF MICROSCOPY, 2003, 211 :19-29
[10]  
THOMPSON NL, 1982, BIOPHYS J, V37, pA355