A miniature and mobile intermittent pneumatic compression device for the prevention of deep-vein thrombosis after joint replacement

被引:17
作者
Ben-Galim, P [1 ]
Steinberg, EL [1 ]
Rosenblatt, Y [1 ]
Parnes, N [1 ]
Menahem, A [1 ]
Arbel, R [1 ]
机构
[1] Tel Aviv Univ, Sackler Fac Med, Tel Aviv Sourasky Med Ctr, Dept Orthoped Surg B, IL-69978 Tel Aviv, Israel
来源
ACTA ORTHOPAEDICA SCANDINAVICA | 2004年 / 75卷 / 05期
关键词
D O I
10.1080/00016470410001466
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The WizAir-DVT is a miniature, lightweight (690 g), battery-operated and mobile intermittent pneumatic compression device (ICD), which enables continuous intraoperative use and immediate patient mobilization postoperatively. We compared its efficacy with a commonly used ICD, the Kendall SCD. Peak femoral vein flow velocity was measured in 20 apparently healthy volunteers at rest and with each device: we found no significant differences between them. A second prospective, randomized, clinical trial was used to compare the efficiency of the device in preventing deep venous thrombosis (DVT) after joint replacement in 50 patients (n = 25/group). None developed DVT. Doppler ultrasonography revealed no significant differences. The WizAir-DVT antithrombotic compression device is as safe and effective as the Kendall SCD.
引用
收藏
页码:584 / 587
页数:4
相关论文
共 11 条
[1]   The fibrinolytic effects of intermittent pneumatic compression - Mechanism of enhanced fibrinolysis [J].
Comerota, AJ ;
Chouhan, V ;
Harada, RN ;
Sun, L ;
Hosking, J ;
Veermansunemi, R ;
Comerota, AJ ;
Schlappy, D ;
Rao, AK .
ANNALS OF SURGERY, 1997, 226 (03) :306-313
[2]  
Cornwell EE, 2002, AM SURGEON, V68, P470
[3]   The effects of external compression on venous blood flow and tissue deformation in the lower leg [J].
Dai, GH ;
Gertler, JP ;
Kamm, RD .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1999, 121 (06) :557-564
[4]   Optimum intermittent pneumatic compression stimulus for lower-limb venous emptying [J].
Delis, KT ;
Azizi, ZA ;
Stevens, RJG ;
Wolfe, JHN ;
Nicolaides, AN .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2000, 19 (03) :261-269
[5]   Randomized clinical trial of intermittent pneumatic compression and low molecular weight heparin in trauma [J].
Ginzburg, E ;
Cohn, SM ;
Lopez, J ;
Jackowski, J ;
Brown, M ;
Hameed, SM .
BRITISH JOURNAL OF SURGERY, 2003, 90 (11) :1338-1344
[6]   EFFECTS OF INTERMITTENT PNEUMATIC LEG COMPRESSION FOR PREVENTION OF POSTOPERATIVE DEEP VENOUS THROMBOSIS WITH SPECIAL REFERENCE TO FIBRINOLYTIC-ACTIVITY [J].
INADA, K ;
KOIKE, S ;
SHIRAI, N ;
MATSUMOTO, K ;
HIROSE, M .
AMERICAN JOURNAL OF SURGERY, 1988, 155 (04) :602-605
[7]   Surgical strategy for management of deep venous thrombosis of the lower extremities [J].
Meissner, AJ ;
Huszcza, S .
WORLD JOURNAL OF SURGERY, 1996, 20 (09) :1149-1155
[8]   Risk of deep-venous thrombosis after hospital discharge in patients having undergone total hip replacement: Double-blind randomised comparison of enoxaparin versus placebo [J].
Planes, A ;
Vochelle, N ;
Darmon, JY ;
Fagola, M ;
Bellaud, M ;
Huet, Y .
LANCET, 1996, 348 (9022) :224-228
[9]  
Salvati EA, 2000, J BONE JOINT SURG AM, V82A, P252
[10]   PROPHYLAXIS AGAINST DEEP-VEIN THROMBOSIS IN TOTAL HIP-REPLACEMENT - COMPARISON OF HEPARIN AND FOOT IMPULSE PUMP [J].
SANTORI, FS ;
VITULLO, A ;
STOPPONI, M ;
SANTORI, N ;
GHERA, S .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (04) :579-583