Efficacy of Quantec rotary instruments for gutta-percha removal

被引:45
作者
Bramante, CM [1 ]
Betti, LV [1 ]
机构
[1] Univ Sao Paulo, Bauru Sch Dent, Dept Endodont, BR-05508 Sao Paulo, Brazil
关键词
gutta-percha removal; nickel-titanium; Quantec system; root canal retreatment; rotary instruments;
D O I
10.1046/j.1365-2591.2000.00340.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim The purpose of this study was to evaluate Quantec SC rotary instruments for removal of gutta-percha during retreatment of straight root canals. Methodology The root canals of 30 central incisors were instrumented and obturated before the teeth were randomly divided into three groups of 10 specimens each. Quantec SC rotary instruments and a 16:1 reduction handpiece powered by an electric motor were then used to remove the gutta-percha and sealer from canals. A different speed was used in each group (group 1, 350 r.p.m,; group 2, 700 r.p.m.: and group 3, 1500 r.p.m,), The followings factors were evaluated: time taken to reach working Length, time for gutta-percha removal, total time, apically extruded material during filling removal and number of fractured instruments. Radiographs were taken after the filling removal and after the canal wall cleanliness was evaluated. The teeth were grooved longitudinally, divided, and the walls of each half were evaluated visually for cleanliness. They were then digitized using a scanner and the residual debris measured. Results The group in which a speed of 1500 r.p.m. was used had significantly faster treatment than the other groups. The amount of apically extruded material was not significantly different between groups. The only significant difference between groups for canal cleanliness was the middle third by radiographic evaluation: the group of 350 r.p.m. had larger amount of debris than the others. In group 1, six instruments fractured; in group 2, four instruments fractured; and in group 3, one instrument fractured. Conclusions Cleanliness and residual debris were equivalent for each group, but the use of 1500 r.p.m. speed was more rapid and fewer instruments fractured.
引用
收藏
页码:463 / 467
页数:5
相关论文
共 13 条
[1]  
BERGENHOLTZ G, 1979, J ENDODONT, V5, P340
[2]   A comparison of stainless steel Flexofiles and nickel-titanium NiTiFlex files during the shaping of simulated canals [J].
Bishop, K ;
Dummer, PMH .
INTERNATIONAL ENDODONTIC JOURNAL, 1997, 30 (01) :25-34
[3]   A COMPARISON OF CANAL PREPARATION WITH NICKEL-TITANIUM AND STAINLESS-STEEL INSTRUMENTS [J].
ESPOSITO, PT ;
CUNNINGHAM, CJ .
JOURNAL OF ENDODONTICS, 1995, 21 (04) :173-176
[4]   ENDODONTIC RETREATMENT - CASE SELECTION AND TECHNIQUE .1. CRITERIA FOR CASE SELECTION [J].
FRIEDMAN, S ;
STABHOLZ, A .
JOURNAL OF ENDODONTICS, 1986, 12 (01) :28-35
[5]   ENDODONTIC RETREATMENT - CASE SELECTION AND TECHNIQUE .3. RETREATMENT TECHNIQUES [J].
FRIEDMAN, S ;
STABHOLZ, A ;
TAMSE, A .
JOURNAL OF ENDODONTICS, 1990, 16 (11) :543-549
[6]   A COMPARISON OF ROOT-CANAL PREPARATIONS USING NI-TI HAND, NI-TI ENGINE-DRIVEN, AND K-FLEX ENDODONTIC INSTRUMENTS [J].
GLOSSON, CR ;
HALLER, RH ;
DOVE, SB ;
DELRIO, CE .
JOURNAL OF ENDODONTICS, 1995, 21 (03) :146-151
[7]  
Hulsmann M, 1997, INT ENDOD J, V30, P227
[8]   EFFECTIVENESS OF HALOTHANE USED WITH ULTRASONIC OR HAND INSTRUMENTATION TO REMOVE GUTTA-PERCHA FROM THE ROOT-CANAL [J].
LADLEY, RW ;
CAMPBELL, AD ;
HICKS, ML ;
LI, SH .
JOURNAL OF ENDODONTICS, 1991, 17 (05) :221-224
[9]  
MCSPADDEN JT, 1996, ADV GEOMETRIES ENDOD
[10]   ENDODONTIC RETREATMENT - CASE SELECTION AND TECHNIQUE .2. TREATMENT PLANNING FOR RETREATMENT [J].
STABHOLZ, A ;