Evaluation of a commercial serological kit for detection of salivary immunoglobulin G to Helicobacter pylori:: a multicentre study

被引:26
作者
Luzza, F [1 ]
Imeneo, M [1 ]
Marasco, A [1 ]
Crotta, S [1 ]
Ierardi, E [1 ]
Usai, P [1 ]
Virgilio, C [1 ]
Nardone, G [1 ]
Marchi, S [1 ]
Sanna, G [1 ]
Perri, F [1 ]
Zagari, RM [1 ]
Bazzoli, F [1 ]
机构
[1] Univ Catanzaro Magna Graecia, Dipartimento Med Sperimentale & Clin, I-88100 Catanzaro, Italy
关键词
diagnosis; Helicobacter pylori; saliva testing; salivary IgG; serum IgG;
D O I
10.1097/00042737-200012100-00008
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
objective To evaluate the usefulness of a commercial serological enzyme immunoassay (EIA) for detecting salivary immunoglobulin (Ig) G to Helicobacter pylori. Design Prospective, multicentre study. Setting The study was conducted at II gastroenterology hospital centres throughout Italy. Participants and interventions Two hundred and thirteen dyspeptic patients underwent gastroscopy with antral biopsies. At each centre, two of the following three tests for ii pylori diagnosis were performed: urease quick test, histology, and C-13-urea breath test Samples of unstimulated saliva and venous blood were collected from each patient, Salivary and serum H. pylori IgG were determined with the EIA Helori-test IgG (Eurospital, Trieste, Italy), Main outcome measures Sensitivity, specificity, positive and negative predictive values, and accuracy of the salivary and serum EIAs. Results By using a receiver operating characteristic curve, salivary EIA yielded 81% (95% confidence interval 73-87%) sensitivity, 73% (62-83%) specificity, 84% (76-90%) positive predictive value, 69% (58-79%) negative predictive value, and 78% (72-84%) accuracy. At the cutoff suggested by the manufacturer, serum EIA had 90% (84-95%) sensitivity, 78% (67-86%) specificity, 88% (81-93%) positive predictive value, 82% (71-90%) negative predictive value, and 86% (81-91%) accuracy. Conclusion In this large multicentre study, detection of salivary H. pylori IgG with a commercial serological EIA was a fairly accurate diagnostic method. Data confirm that saliva testing does not compare favourably with serology in the assessment of ii pylori status. (C) 2000 Lippincott Williams & Wilkins.
引用
收藏
页码:1117 / 1120
页数:4
相关论文
共 22 条
  • [1] Is saliva serology useful for the diagnosis of Helicobacter pylori?
    Christie, JML
    McNulty, CAM
    Shepherd, NA
    Valori, RM
    [J]. GUT, 1996, 39 (01) : 27 - 30
  • [2] Performance characteristics of an enzyme-linked immunosorbent assay for determining salivary immunoglobulin G response to Helicobacter pylori
    De Pascalis, R
    Del Pezzo, M
    Nardone, G
    Budillon, G
    Lavitola, A
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (02) : 430 - 432
  • [3] *EUR HEL PYL STUD, 1997, GUT, V41, pS10
  • [4] Fallone CA, 1996, AM J GASTROENTEROL, V91, P1145
  • [5] Karvar S, 1998, GUT, V42, P900
  • [6] Evaluation of commercially available Helicobacter pylori serology kits:: a review
    Laheij, RJF
    Straatman, H
    Jansen, JBMJ
    Verbeek, ALM
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1998, 36 (10) : 2803 - 2809
  • [7] Evaluation of salivary antibodies to detect infection with Helicobacter pylori
    Loeb, MB
    Riddell, RH
    James, C
    Hunt, R
    Smaill, FM
    [J]. CANADIAN JOURNAL OF GASTROENTEROLOGY, 1997, 11 (05): : 437 - 440
  • [8] Luzza F, 1998, ITAL J GASTROENTEROL, V30, P36
  • [9] Evaluation of a commercial serological kit for detection of salivary immunoglobulin G to Helicobacter pylori
    Luzza, F
    Imeneo, M
    Maletta, M
    Graziano, P
    Marasco, R
    Monteleone, G
    Caradonna, M
    Pallone, F
    [J]. EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 1997, 16 (10) : 775 - 777
  • [10] LUZZA F, 1995, FEMS IMMUNOL MED MIC, V10, P281, DOI 10.1111/j.1574-695X.1995.tb00044.x