Three-dimensional hydrodynamic focusing in two-layer polydimethylsiloxane (PDMS) microchannels

被引:111
作者
Chang, Chih-Chang [1 ]
Huang, Zhi-Xiong [1 ]
Yang, Ruey-Jen [1 ]
机构
[1] Natl Cheng Kung Univ, Dept Engn Sci, Tainan 70101, Taiwan
关键词
D O I
10.1088/0960-1317/17/8/009
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this work, we designed and fabricated a three-dimensional hydrodynamic focusing microfluidic device. The device comprises a two-layer PDMS microchannel structure. There are four inlet ports and one outlet port. The fluids are all injected by syringe pumps. A sample flow stream was first vertically constrained into a narrow stream, and then horizontally focused on one small core region from a cross-section perspective, which is useful for cell/particle counting. We showed the numerical and experimental images of the focused stream shape from a cross-section perspective; experimental images were captured using a confocal fluorescence microscope. We also investigated the effect of channel aspect ratio on the vertical focusing effect using CFD simulations. The results showed that the sample flow can be focused successfully in the lower aspect ratio of the main channel (slightly greater than 0.5) in our design. Furthermore, the effect of the Reynolds number on the vertical focusing effect was also investigated. The numerical results showed that the rectangular-like shape of the focused stream from the cross-section perspective was deformed as the Reynolds number was high due to stronger secondary flows produced in the vertical focusing unit. This phenomenon was also demonstrated experimentally. The device only works well at low Reynolds numbers (approximately less than 5). The device can be integrated into an on-chip flow cytometer.
引用
收藏
页码:1479 / 1486
页数:8
相关论文
共 38 条
[1]   Formation of dispersions using "flow focusing" in microchannels [J].
Anna, SL ;
Bontoux, N ;
Stone, HA .
APPLIED PHYSICS LETTERS, 2003, 82 (03) :364-366
[2]  
[Anonymous], FLOW CYTOMETRY INSTR
[3]   A numerical method for solving incompressible viscous flow problems (Reprinted from the Journal of Computational Physics, vol 2, pg 12-26, 1997) [J].
Chorin, AJ .
JOURNAL OF COMPUTATIONAL PHYSICS, 1997, 135 (02) :118-125
[4]  
Chung S, 2003, MICROSYST TECHNOL, V9, P525, DOI [10.1007/s00542-003-0302-2, 10.1007/S00542-003-0302-2]
[5]  
CROSLANDTAYLOR P, 1953, NATURE, V17, P37
[6]   Micro total analysis systems. Latest advancements and trends [J].
Dittrich, Petra S. ;
Tachikawa, Kaoru ;
Manz, Andreas .
ANALYTICAL CHEMISTRY, 2006, 78 (12) :3887-3907
[7]   Rapid prototyping of microfluidic systems in poly(dimethylsiloxane) [J].
Duffy, DC ;
McDonald, JC ;
Schueller, OJA ;
Whitesides, GM .
ANALYTICAL CHEMISTRY, 1998, 70 (23) :4974-4984
[8]   Electrokinetically driven micro flow cytometers with integrated fiber optics for on-line cell/particle detection [J].
Fu, LM ;
Yang, RJ ;
Lin, CH ;
Pan, YJ ;
Lee, GB .
ANALYTICA CHIMICA ACTA, 2004, 507 (01) :163-169
[9]   Multiple injection techniques for microfluidic sample handling [J].
Fu, LM ;
Yang, RJ ;
Lee, GB ;
Pan, YJ .
ELECTROPHORESIS, 2003, 24 (17) :3026-3032
[10]   Perfectly monodisperse microbubbling by capillary flow focusing -: art. no. 274501 [J].
Gañán-Calvo, AM ;
Gordillo, JM .
PHYSICAL REVIEW LETTERS, 2001, 87 (27)