A prospective multicenter study on the safety of a pacemaker with automatic energy control:: Influence of the electrical factor on chronic stimulation threshold.

被引:14
作者
Madrid, AH
Olagüe, J
Cercas, A
Del Ojo, JL
Muñoz, F
Moro, C
Sanz, O
机构
[1] Hosp Ramon & Cajal, Arrhythmia Unit, E-28034 Madrid, Spain
[2] Univ Alcala de Henares, Madrid, Spain
[3] Hosp Univ La Fe, Valencia, Spain
[4] Hosp Seguridad Social, Jerez de la Frontera, Spain
[5] Hosp Univ Valme, Seville, Spain
[6] Hosp Rio Hortega, Valladolid, Spain
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2000年 / 23卷 / 09期
关键词
cardiac pacing; electrodes; pacemaker;
D O I
10.1111/j.1540-8159.2000.tb00963.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The effectiveness and safety of a pacemaker with automatic control of capture was evaluated in 162 patients followed at 27 Spanish centers. The aim of our prospective, multicenter, and randomized trial was to determine the relationship between the voltage output of the pulse generator and the stimulation threshold. We randomized 162 patients (107 men, mean age 75 +/- 12 years). We implanted a ventricular pacemaker model Regency SR+ or SC+ with Pacesetter's low polarization bipolar leads Membrane E 1450. The patients were randomized to receive Autocapture or not; group I (81 patients) Autocapture On, pulse output automatically adjusted and group II (81 patients) Autocapture Off, fixed output parameters (3.9 V, 0.37 ms). We performed a 6-month follow-up measuring stimulation threshold by means of the VARIO test and Autocapture test, evoked response signal, and R wave signal. The mean R wave was 15.77 +/- 3.5 mV at the end of the follow-up for group I; and 14.91 +/- 6.8 mV for group II (P = NS). The measured evoked response at the end of the follow-up was 9.25 mV in Group I and 8.48 mV in Group II (P = NS). The stimulation threshold was not different between groups. The current density created with the voltage and pulse width used in this study (less than or equal to 3.9 V and 0.37 ms) at the tip of this electrode during the maturation process had no influence on the development of the chronic detection and stimulation thresholds.
引用
收藏
页码:1359 / 1364
页数:6
相关论文
共 19 条
[1]   CHARACTERIZATION OF CHRONICALLY IMPLANTED VENTRICULAR ENDOCARDIAL PACING LEADS [J].
ANGELLO, DA ;
MCANULTY, JH ;
DOBBS, J .
AMERICAN HEART JOURNAL, 1984, 107 (06) :1142-1145
[2]   TECHNIQUES AND SIGNIFICANCE OF THRESHOLD MEASUREMENT FOR CARDIAC PACING - RELATIONSHIP TO OUTPUT CIRCUIT OF CARDIAC-PACEMAKERS [J].
BAROLD, SS ;
WINNER, JA .
CHEST, 1976, 70 (06) :760-766
[3]   COMPARISON OF A POROUS-SURFACED WITH A TOTALLY POROUS VENTRICULAR ENDOCARDIAL PACING ELECTRODE [J].
BOBYN, JD ;
WILSON, GJ ;
MYCYK, TR ;
KLEMENT, P ;
TAIT, GA ;
PILLIAR, RM ;
MACGREGOR, DC .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1981, 4 (04) :405-416
[4]   DECREASING ELECTRODE SIZE AND INCREASING EFFICIENCY OF CARDIAC STIMULATION [J].
FURMAN, S ;
PARKER, B ;
ESCHER, DJW .
JOURNAL OF SURGICAL RESEARCH, 1971, 11 (03) :105-&
[5]  
HENSON PM, 1980, AM J PATHOL, V101, P494
[6]   ACUTE AND CHRONIC CYCLE LENGTH DEPENDENT INCREASE IN VENTRICULAR PACING THRESHOLD [J].
HOOK, BG ;
PERLMAN, RL ;
CALLANS, DJ ;
HANNA, MS ;
KLEIMAN, RB ;
FLORES, BT ;
MARCHLINSKI, FE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (10) :1437-1444
[7]   EFFECTS OF ANTIARRHYTHMIC DRUGS ON THE CHRONIC PACING THRESHOLD AND THE ENDOCARDIAL R-WAVE AMPLITUDE IN THE CONSCIOUS DOG [J].
HUANG, SK ;
HEDBERG, PS ;
MARCUS, FI .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (05) :660-669
[8]   COMPARISON OF PACING ELECTRODES OF DIFFERENT SHAPE AND MATERIAL - RECOMMENDATIONS [J].
IRNICH, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1983, 6 (02) :422-426
[9]   A NEW LOW THRESHOLD PLATINIZED EPICARDIAL PACING ELECTRODE - COMPARATIVE-EVALUATION IN IMMATURE CANINES [J].
KARPAWICH, PP ;
STOKES, KB ;
HELLAND, JR ;
JUSTICE, CD ;
ROSKAMP, JO .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (08) :1139-1148
[10]  
Liebold A, 1997, Z KARDIOL, V86, P521, DOI 10.1007/s003920050088