Use of rapid prototyping models in the planning of percutaneous pulmonary valved stent implantation

被引:54
作者
Armillotta, A.
Bonhoeffer, P.
Dubini, G.
Ferragina, S.
Migliavacca, F.
Sala, G.
Schievano, S.
机构
[1] Politecn Milan, Dipartimento Meccan, I-20156 Milan, Italy
[2] Great Ormond St Hosp Sick Children, Cardiothorac Unit, London, England
[3] Politecn Milan, Dipartimento Ingn Strutturale, Lab Biol Struct Mech, I-20133 Milan, Italy
[4] Politecn Milan, Dipartimento Ingn Aerospaziale, I-20156 Milan, Italy
关键词
vascular models; rapid prototyping; fused deposition modelling; vacuum; casting; stent implantation;
D O I
10.1243/09544119JEIM83
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Percutaneous replacement of the pulmonary valve is a recently developed interventional technique which involves the implantation of a valved stent in the pulmonary trunk. It relies upon careful consideration of patient anatomy for both stent design and detailed procedure planning. Medical imaging data in the form of two-dimensional scans and three-dimensional interactive graphics offer only limited support for these tasks. The paper reports the results of an experimental investigation on the use of arterial models built by rapid proto-typing techniques. An analysis of clinical needs has helped to specify proper requirements for such model properties as cost, strength, accuracy, elastic compliance, and optical transparency. Two different process chains, based on the fused deposition modelling technique and on the vacuum casting of thermoset resins in rubber moulds, have been tested for prototype fabrication. The use of anatomical models has allowed the cardiologist's confidence in patient selection, prosthesis fabrication, and final implantation to be significantly improved.
引用
收藏
页码:407 / 416
页数:10
相关论文
共 26 条
[1]   Experimental flow studies in exact-replica phantoms of atherosclerotic carotid bifurcations under steady input conditions [J].
Bale-Glickman, J ;
Selby, K ;
Saloner, D ;
Savas, Ö .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2003, 125 (01) :38-48
[2]   Percutaneous insertion of the pulmonary valve [J].
Bonhoeffer, P ;
Boudjemline, Y ;
Qureshi, SA ;
Le Bidois, J ;
Iserin, L ;
Acar, P ;
Merckx, J ;
Kachaner, J ;
Sidi, D .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (10) :1664-1669
[3]   Percutaneous pulmonary valve replacement in a large right ventricular outflow tract - An experimental study [J].
Boudjemline, Y ;
Agnoletti, G ;
Bonnet, D ;
Sidi, D ;
Bonhoeffer, P .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (06) :1082-1087
[4]   Percutaneous aortic valve replacement in animals [J].
Boudjemline, Y ;
Bonhoeffer, P .
CIRCULATION, 2004, 109 (10) :E161-E161
[5]   Steps toward percutaneous aortic valve replacement [J].
Boudjemline, Y ;
Bonhoeffer, P .
CIRCULATION, 2002, 105 (06) :775-778
[6]  
Boudjemline Y, 2001, HEART, V86, P705
[7]  
BOUDJEMLINE Y, 2002, MED SCI MONITOR, V8
[8]   RIGHT-VENTRICULAR OUTFLOW RECONSTRUCTION WITH CRYOPRESERVED HOMOGRAFTS IN PEDIATRIC-PATIENTS - INTERMEDIATE-TERM FOLLOW-UP WITH SERIAL ECHOCARDIOGRAPHIC ASSESSMENT [J].
CHAN, KC ;
FYFE, DA ;
MCKAY, CA ;
SADE, RM ;
CRAWFORD, FA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (02) :483-489
[9]   Computer aided design and fabrication of models for in vitro studies of vascular fluid dynamics [J].
Chong, CK ;
Rowe, CS ;
Sivanesan, S ;
Rattray, A ;
Black, RA ;
Shortland, AP ;
How, TV .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1999, 213 (H1) :1-4
[10]   Single-step stereolithography of complex anatomical models for optical flow measurements [J].
de Zélicourt, D ;
Pekkan, K ;
Kitajima, H ;
Frakes, D ;
Yoganathan, AP .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (01) :204-207