pO2 measurements by phosphorescence quenching:: characteristics and applications of an automated system

被引:63
作者
Kerger, H [1 ]
Groth, G
Kalenka, A
Vajkoczy, P
Tsai, AG
Intaglietta, M
机构
[1] Univ Hosp Mannheim, Dept Operat Crit Care Med, Dept Anesthesiol, D-68135 Mannheim, Germany
[2] Univ Hosp Mannheim, Dept Neurosurg, D-68135 Mannheim, Germany
[3] Univ Calif San Diego, Dept Bioengn, La Jolla, CA 92093 USA
关键词
pO(2); measurement microcirculation; phosphorescence quenching;
D O I
10.1016/S0026-2862(02)00027-4
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
An automated system for pO(2) analysis based upon phosphorescence quenching was tested. The system was calibrated in vitro with capillary samples of saline and blood. Results were compared to a conventional measuring procedure wherein pO(2) was calculated off-line by computer fitting of phosphorescence decay signals. PO2 measurements obtained by the automated system were correlated (r(2) = 0.98) with readings simultaneously generated by a blood gas analyzer, accuracy being highest in the low (0-20 mm Hg) and medium pO(2) ranges (21-70 mm He). Measurements in in vivo studies in the hamster skin-fold preparation were similar to previously reported results. The automated system fits the phosphorescence decay data to a single exponential and allows repeated pO(2) measurements in rapid sequence. (C) 2003 Elsevier Science (USA). All rights reserved.
引用
收藏
页码:32 / 38
页数:7
相关论文
共 22 条
[1]   In vivo pO2 measurements in hamster skinfold by recessed pO2 microelectrodes and phosphorescence quenching are in agreement [J].
Buerk, DG ;
Tsai, AG ;
Intaglietta, M ;
Johnson, PC .
MICROCIRCULATION, 1998, 5 (2-3) :219-225
[2]   TECHNICAL REPORT - A NEW CHAMBER TECHNIQUE FOR MICRO-VASCULAR STUDIES IN UNANESTHETIZED HAMSTERS [J].
ENDRICH, B ;
ASAISHI, K ;
GOTZ, A ;
MESSMER, K .
RESEARCH IN EXPERIMENTAL MEDICINE, 1980, 177 (02) :125-134
[3]   Analysis of phosphorescence in heterogeneous systems using distributions of quencher concentration [J].
Golub, AS ;
Popel, AS ;
Zheng, L ;
Pittman, RN .
BIOPHYSICAL JOURNAL, 1997, 73 (01) :452-465
[4]   Phosphorescence quenching method for measurement of intracellular Po2 in isolated skeletal muscle fibers [J].
Hogan, MC .
JOURNAL OF APPLIED PHYSIOLOGY, 1999, 86 (02) :720-724
[5]   MICROVASCULAR MEASUREMENTS BY VIDEO IMAGE SHEARING AND SPLITTING [J].
INTAGLIETTA, M ;
TOMPKINS, WR .
MICROVASCULAR RESEARCH, 1973, 5 (03) :309-312
[6]   Microvascular oxygen delivery and interstitial oxygenation during sodium pentobarbital anesthesia [J].
Kerger, H ;
Saltzman, DJ ;
Gonzales, A ;
Tsai, AG ;
vanAckern, K ;
Winslow, RM ;
Intaglietta, M .
ANESTHESIOLOGY, 1997, 86 (02) :372-386
[7]   Systemic and microcirculatory effects of autologous whole blood resuscitation in severe hemorrhagic shock [J].
Kerger, H ;
Waschke, KF ;
Ackern, KV ;
Tsai, AG ;
Intaglietta, M .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1999, 276 (06) :H2035-H2043
[8]   SYSTEMIC AND SUBCUTANEOUS MICROVASCULAR OXYGEN-TENSION IN CONSCIOUS SYRIAN GOLDEN-HAMSTERS [J].
KERGER, H ;
TORRES, IP ;
RIVAS, M ;
WINSLOW, RM ;
INTAGLIETTA, M .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1995, 268 (02) :H802-H810
[9]   Depression of endothelial and smooth muscle cell oxygen consumption by endotoxin [J].
Motterlini, R ;
Kerger, H ;
Green, CJ ;
Winslow, RM ;
Intaglietta, M .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1998, 275 (03) :H776-H782
[10]  
Pawlowski Marek, 1992, V316, P179