Volume CT with a flat-panel detector on a mobile, isocentric C-arm: Pre-clinical investigation in guidance of minimally invasive surgery

被引:255
作者
Siewerdsen, JH [1 ]
Moseley, DJ
Burch, S
Bisland, SK
Bogaards, A
Wilson, BC
Jaffray, DA
机构
[1] Princess Margaret Hosp, Ontario Canc Inst, Toronto, ON M5G 2M9, Canada
[2] Univ Toronto, Dept Med Biophys, Toronto, ON M5G 2M9, Canada
[3] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5G 2M9, Canada
[4] Princess Margaret Hosp, Radiat Med Program, Toronto, ON M5G 2M9, Canada
关键词
image-guided surgery; minimally invasive; volume CT; cone-beam CT; fluoroscopy; flat-panel detector; C-arm; image quality; spine surgery; spinal metastases; photodynamic therapy;
D O I
10.1118/1.1836331
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A mobile isocentric C-arm (Siemens PowerMobil) has been modified in our laboratory to include a large area flat-panel detector (in place of the x-ray image intensifier), providing multi-mode fluoroscopy and cone-beam computed tomography (CT) imaging capability. This platform represents a promising technology for minimally invasive, image-guided surgical procedures where precision in the placement of interventional tools with respect to bony and soft-tissue structures is critical. The image quality and performance in surgical guidance was investigated in pre-clinical evaluation in image-guided spinal surgery. The control, acquisition, and reconstruction system are described. The reproducibility of geometric calibration, essential to achieving high three-dimensional (3D) image quality, is tested over extended time scales (7 months) and across a broad range in C-arm, angulation (up to 45degrees), quantifying the effect of improper calibration on spatial resolution, soft-tissue visibility, and image artifacts. Phantom studies were performed to investigate the precision of 3D localization (viz., fiber optic probes within a vertebral body) and effect of lateral projection truncation (limited field of view) on soft-tissue detectability in image reconstructions. Pre-clinical investigation was undertaken in a specific spinal procedure (photodynamic therapy of spinal metastases) in five animal subjects (pigs). In each procedure, placement of fiber optic catheters in two vertebrae (L1 and L2) was guided by fluoroscopy and cone-beam CT. Experience across five procedures is reported, focusing on 3D image quality, the effects of respiratory motion, limited field of view, reconstruction filter, and imaging dose. Overall, the intraoperative cone-beam CT images were sufficient for guidance of needles and catheters with respect to bony anatomy and improved surgical performance and confidence through 3D visualization and verification of transpedicular trajectories and tool placement. Future investigation includes improvement in image quality, particularly regarding x-ray scatter, motion artifacts and field of view, and integration with optical tracking and navigation systems. (C) 2005 American Association of Physicists in Medicine.
引用
收藏
页码:241 / 254
页数:14
相关论文
共 58 条
[1]   DEMONSTRATION OF MEGAVOLTAGE AND DIAGNOSTIC-X-RAY IMAGING WITH HYDROGENATED AMORPHOUS-SILICON ARRAYS [J].
ANTONUK, LE ;
BOUDRY, J ;
HUANG, WD ;
MCSHAN, DL ;
MORTON, EJ ;
YORKSTON, J ;
LONGO, MJ ;
STREET, RA .
MEDICAL PHYSICS, 1992, 19 (06) :1455-1466
[2]   Development and implementation of intraoperative magnetic resonance imaging and its neurosurgical applications [J].
Black, PM ;
Moriarty, T ;
Alexander, E ;
Stieg, P ;
Woodard, EJ ;
Gleason, PL ;
Martin, CH ;
Kikinis, R ;
Schwartz, RB ;
Jolesz, FA .
NEUROSURGERY, 1997, 41 (04) :831-842
[3]   MONTE-CARLO SIMULATION OF THE SCATTERED RADIATION DISTRIBUTION IN DIAGNOSTIC-RADIOLOGY [J].
BOONE, JM ;
SEIBERT, JA .
MEDICAL PHYSICS, 1988, 15 (05) :713-720
[4]   Dedicated breast CT: Radiation dose and image quality evaluation [J].
Boone, JM ;
Nelson, TR ;
Lindfors, KK ;
Seibert, JA .
RADIOLOGY, 2001, 221 (03) :657-667
[5]   AN ANALYTICAL MODEL OF THE SCATTERED RADIATION DISTRIBUTION IN DIAGNOSTIC-RADIOLOGY [J].
BOONE, JM ;
SEIBERT, JA .
MEDICAL PHYSICS, 1988, 15 (05) :721-725
[6]  
BURCH S, IN PRESS J SPINE
[7]  
BURCH S, UNPUB J BIOMED OPT
[8]   Cone-beam volume CT breast imaging: Feasibility study [J].
Chen, B ;
Ning, R .
MEDICAL PHYSICS, 2002, 29 (05) :755-770
[9]   40 x 30 cm flat panel imager for angiography, R&F, and cone-beam CT applications [J].
Colbeth, RE ;
Boyce, S ;
Fong, R ;
Gray, K ;
Harris, R ;
Job, ID ;
Mollov, I ;
Nepo, B ;
Pavkovich, J ;
Taie-Nobarie, N ;
Seppi, EJ ;
Shapiro, EG ;
Wright, MD ;
Webb, C ;
Yu, JAM .
MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 :94-102
[10]   3-DIMENSIONAL IMAGE-RECONSTRUCTION FROM COMPLETE PROJECTIONS [J].
DEFRISE, M ;
TOWNSEND, DW ;
CLACK, R .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (05) :573-587