Design for a simplified cochlear implant system

被引:40
作者
An, Soon Kwan
Park, Se-Ik
Jun, Sang Beom
Lee, Choong Jae
Byun, Kyung Min
Sung, Jung Hyun
Wilson, Blake S.
Rebscher, Stephen J.
Oh, Seung Ha
Kim, Sung June [1 ]
机构
[1] Seoul Natl Univ, Sch Elect Engn & Comp Sci, Seoul 151744, South Korea
[2] Seoul Natl Univ, Nano Bioelect & Syst Res Ctr, Seoul 151744, South Korea
[3] Nurobiosys Corp, Seoul 151050, South Korea
[4] Duke Univ, Med Ctr, Dept Surg, Durham, NC 27710 USA
[5] Univ Calif San Francisco, Epstein Lab, Dept Otolaryngol, San Francisco, CA 93143 USA
[6] Seoul Natl Univ, Dept Otolaryngol, Seoul 110744, South Korea
关键词
auditory prosthesis; cochlear implant system; continuous interleaved sampling; electrode array; hermetic package; neural prosthesis; neural stimulation;
D O I
10.1109/TBME.2007.895372
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A simplified cochlear implant (CI) system would be appropriate for widespread use in developing countries. Here, we describe a Cl that we have designed to realize such a concept. The system implements 8 channels of processing and stimulation using the continuous interleaved sampling (CIS) strategy. A generic digital signal processing (DSP) chip is used for the processing, and the filtering functions are performed with a fast Fourier transform (FFT) of a microphone or other input. Data derived from the processing are transmitted through an inductive link using pulse width modulation (PWM) encoding and amplitude shift keying (ASK) modulation. The same link is used in the reverse direction for backward telemetry of electrode and system information. A custom receiver-stimulator chip has been developed that demodulates incoming data using pulse counting and produces charge balanced biphasic pulses at 1000 pulses/s/electrode. This chip is encased in a titanium package that is hermetically sealed using a simple but effective method. A low cost metal-silicon hybrid mold has been developed for fabricating an intracochlear electrode array with 16 ball-shaped stimulating contacts.
引用
收藏
页码:973 / 982
页数:10
相关论文
共 25 条
[1]  
[Anonymous], 1910, B BUR FISH
[2]  
Brill SM, 1997, AM J OTOL, V18, pS104
[3]   MULTIPLE-ELECTRODE ARRAY FOR A COCHLEAR IMPLANT [J].
CLARK, GM ;
HALLWORTH, RJ .
JOURNAL OF LARYNGOLOGY AND OTOLOGY, 1976, 90 (07) :623-627
[4]   Speech intelligibility as a function of the number of channels of stimulation for signal processors using sine-wave and noise-band outputs [J].
Dorman, MF ;
Loizou, PC ;
Rainey, D .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1997, 102 (04) :2403-2411
[5]   A high-efficiency power and data transmission system for biomedical implanted electronic devices [J].
Hamici, Z ;
Itti, R ;
Champier, J .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1996, 7 (02) :192-201
[6]  
HAMIDA AB, 1998, P 24 ANN C IEEE IND, V3, P1388
[7]   COLLECTOR AMPLITUDE-MODULATION OF THE CLASS-E TUNED POWER-AMPLIFIER [J].
KAZIMIERCZUK, M .
IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS, 1984, 31 (06) :543-549
[8]  
Kessler D. K., 1995, Annals of Otology Rhinology and Laryngology, V104, P283
[9]   Three-dimensional analysis of electrode behavior in a human cochlear model [J].
Lim, YS ;
Park, SI ;
Kim, YH ;
Oh, SH ;
Kim, SJ .
MEDICAL ENGINEERING & PHYSICS, 2005, 27 (08) :695-703
[10]  
LOEB GE, 1994, P INT COCHL IMPL SPE, P24