Simulation and experimental observation of contact conditions between stents and artery models

被引:54
作者
Takashima, Kazuto
Kitou, Takayuki
Mori, Koji
Ikeuchi, Ken
机构
[1] Kyoto Univ, Inst Frontier Med Sci, Dept Nano Biomech, Kyoto 6068507, Japan
[2] Yamaguchi Univ, Grad Sch Med, Ube, Yamaguchi 7558611, Japan
关键词
stent; artery; finite element method; in vitro testing; stress distribution;
D O I
10.1016/j.medengphy.2006.04.003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Treatment of coronary artery stenosis with percutaneous coronary angioplasty and stenting is sometimes complicated by neointimal hyperplasia, possibly due to interaction of the stent with the arterial wall within a specific contact area. Therefore, we characterized the stress distribution at the contacts between the stent and the artery using mathematical and experimental modeling (an arterial cylinder model with a tube-like structure and an arterial stenosis model, consisting of a tube and plaque portion) and two kinds of link stents with different numbers of cells and links. First, the contact condition was investigated using a finite element method (FEM). Second. experimental visualization of the contact area between the stent and the artery models was performed. Comparison of the experimental results with the FEM analysis revealed that the contact area between the stent (with a high number of cells and links) and the artery model was distributed over the total surface of the stent. Further, values obtained from the experimental distribution and the calculated distribution were similar. These data indicate that experimental modeling and FEM analysis are useful methods for analyzing the relationship between stent structure and stent/wall stress distributions and may help guide the design of new stents. (c) 2006 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:326 / 335
页数:10
相关论文
共 31 条
[1]  
Carter AJ, 2000, CATHETER CARDIO INTE, V51, P112, DOI 10.1002/1522-726X(200009)51:1<112::AID-CCD26>3.0.CO
[2]  
2-H
[3]   Coronary stenting with a novel stainless steel balloon-expandable stent: Determinants of neointimal formation and changes in arterial geometry after placement in an atherosclerotic model [J].
Carter, AJ ;
Laird, JR ;
Kufs, WM ;
Bailey, L ;
Hoopes, TG ;
Reeves, T ;
Farb, A ;
Virmani, R .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 27 (05) :1270-1277
[4]  
David ChuaS N., 2004, J MATER PROCESS TE 2, V155-156, P1772
[5]   A RANDOMIZED COMPARISON OF CORONARY-STENT PLACEMENT AND BALLOON ANGIOPLASTY IN THE TREATMENT OF CORONARY-ARTERY DISEASE [J].
FISCHMAN, DL ;
LEON, MB ;
BAIM, DS ;
SCHATZ, RA ;
SAVAGE, MP ;
PENN, I ;
DETRE, K ;
VELTRI, L ;
RICCI, D ;
NOBUYOSHI, M ;
CLEMAN, M ;
HEUSER, R ;
ALMOND, D ;
TEIRSTEIN, PS ;
FISH, RD ;
COLOMBO, A ;
BRINKER, J ;
MOSES, J ;
SHAKNOVICH, A ;
HIRSHFELD, J ;
BAILEY, S ;
ELLIS, S ;
RAKE, R ;
GOLDBERG, S .
NEW ENGLAND JOURNAL OF MEDICINE, 1994, 331 (08) :496-501
[6]   STRENGTH, ELASTICITY, AND PLASTICITY OF EXPANDABLE METAL STENTS - IN-VITRO STUDIES WITH 3 TYPES OF STRESS [J].
FLUECKIGER, F ;
STERNTHAL, H ;
KLEIN, GE ;
ASCHAUER, M ;
SZOLAR, D ;
KLEINHAPPL, G .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1994, 5 (05) :745-750
[7]   Stent and artery geometry determine intimal thickening independent of arterial injury [J].
Garasic, JM ;
Edelman, ER ;
Squire, JC ;
Seifert, P ;
Williams, MS ;
Rogers, C .
CIRCULATION, 2000, 101 (07) :812-818
[8]   Chronic arterial responses to stent implantation: A serial intravascular ultrasound analysis of Palmaz-Schatz stents in native coronary arteries [J].
Hoffmann, R ;
Mintz, GS ;
Popma, JJ ;
Satler, LF ;
Pichard, AD ;
Kent, KM ;
Walsh, C ;
Mackell, P ;
Leon, MB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 28 (05) :1134-1139
[9]   Changes in the mechanical environment of stenotic arteries during interaction with stents: Computational assessment of parametric stent designs [J].
Holzapfel, G ;
Stadler, M ;
Gasser, TC .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (01) :166-180
[10]  
KAMIMURA T, 2002, T JPN SOC MECH ENG-B, V68, P424